www.rossmax.com
Specications
Cautions and maintenance
1. This device should be used with nebulizer only.
2. Place the device and nebulizer on a at, horizontal surface while in use.
3. The device must be in a vertical position while in use.
4. Turn the regulator slowly to avoid possible damage.
5. Do not completely unscrew regulator stem.
6. Do not attempt to disassemble the device.
7. Avoid excessive vibration, physical impact and dropping.
8. Do not store the device in direct sunlight, high temperature or humidity.
9. Keep the device away from static electricity.
10. For the rst use, open the rubber vent-plug on the back side of pressure gauge and
then close it in order to balance the pressure after the transportation.
Note: Do not open the rubber vent-plug when the device is in a vertical position to avoid
leakages of oil material from the pressure gauge.
Power source No need
Operating temperature 0 ~ 54ºC
Storage temperature - 10 ~ 60ºC
Air ow scale 1 ~ 10 L/min
accuracy ± 4% of full scale
Pressure scale 0 ~ 60 PSI
accuracy ± 2% of full scale
Weight 1000 g
Dimension 128 (L)*120(W)*158 (H) mm
Rossmax Neb Tester is a portable and easy-to-use device that allows the user to quickly
check the performance of the pump of a compressor nebulizer. Power source is not
required for Neb Tester during operation and the test result is easy to read and interpret.
Neb Tester consists of an oil pressure gauge, ow meter, air tube and stainless steel stand.
This device is used to simulate the working conditions of the pump when the nebulizer is
in operation with a medication cup, and you are able to check:
1. Maximum (free) air ow
2. Operational air ow at a particular pressure
3. Maximum air pressure
Parts name/location
Introduction
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
Flow meter
Regulator
Pressure
gauge
Air tube
Insert air tube into connectorPlace the device in a
horizontal position
Open the rubber vent-plug
and then close it after
Back view
Installation
Operation procedures
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
psi
kg/cm2
5
10
20 30
40
50
60
0
1
2
3
4
1. Check maximum (free) air ow
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Adjust air ow and set the level of the pressure at 0 PSI by turning the
regulator of ow meter.
Step 6: Check the air ow rate.
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
2. Check operational air ow at a particular pressure
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Set the pressure to a desired level by adjusting the regulator of the ow
meter.
Step 6: Check the air ow rate.
Pressure (PSI) Flow rate (L/min)
NA100 / NB500 (N1) 15.0 ≥ 3.5
NE100 / NF100/ NJ100 (N4) 15.0 ≥ 3.5
NK1000/ NB80 / NF80 (N1) 8.0 ≥ 3.0
NB60 / NI60/ NH60 / NF60 (N1) 8.0 ≥ 2.0
NL100 (N2) 10.5 ≥ 3.5
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
3. Check maximum air pressure
Step 1: Turn the regulator of ow meter anticlockwise around 2~3 circles.
Step 2: Insert the other end of air tube into the air outlet of the nebulizer.
Step 3: Power on the nebulizer.
Step 4: Keep the nebulizer running for 30 seconds to warm up.
Step 5: Set the air ow rate at 0 L/min by turning the regulator of ow meter
clockwise completely.
Step 6: Check the values on the pressure gauge.
Step 7: Turn off the nebulizer.
Step 8: Disconnect the air tube from the nebulizer.
Air
compressor
connector
Rubber
vent-plug


Specyfikacje produktu

Marka: Rossmax
Kategoria: Medyczny
Model: Neb Tester

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Rossmax Neb Tester, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Medyczny Rossmax

Rossmax

Rossmax NH60 Instrukcja

24 Września 2024
Rossmax

Rossmax AM30 Instrukcja

23 Września 2024
Rossmax

Rossmax V5 Instrukcja

21 Września 2024
Rossmax

Rossmax V3 Instrukcja

20 Września 2024
Rossmax

Rossmax EB500 Instrukcja

16 Września 2024
Rossmax

Rossmax NA100 Instrukcja

12 Września 2024
Rossmax

Rossmax NB500 Instrukcja

12 Września 2024
Rossmax

Rossmax NW1 Instrukcja

11 Września 2024
Rossmax

Rossmax PF120C Instrukcja

9 Września 2024

Instrukcje Medyczny

Najnowsze instrukcje dla Medyczny

Gima

Gima OXY 100 Instrukcja

27 Września 2024
Gima

Gima OXY 5 Instrukcja

27 Września 2024
Gima

Gima OXY 50 Instrukcja

27 Września 2024
Gima

Gima OXY 1 Instrukcja

27 Września 2024
Gima

Gima OXY 4 Instrukcja

26 Września 2024
Gima

Gima OXY-PED Instrukcja

26 Września 2024
Dittmann

Dittmann ZBE 339 Instrukcja

24 Września 2024