1.0 Object
1.1
To test the performance of Rossmax NE100 nebulizer
2.0 Equipment List
2.1 Rossmax NE100 Nebulizer*2
2.2 Rossmax Nebulizer kit*2
2.3 Malvern Spraytec particle size analyer
2.4 Marple 298 Cascade Impactor
2.5 Chroma 61602 Programmable AC Source
2.6 Shimadzu AUW120D microbalance
2.7 A.P. Buck, Inc. Libra Plus LP-5 sampling pump
2.8 SSI P51-6BarS-A-MD-20mA pressure meter
2.9 Golden Mountain Enterprise Co. Ltd. F33L0096 flow meter
2.10 Humidity/Temperature Meter
2.11 Taiwan Biotech Co., Ltd 0.9% Saline solution
2.12 2.5% NaF solution
2.13 Atrovent Ipratropium Bromide
2.14
Atrovent Flixotide
2.15 AstraZeneca Terbutaline Sulphate
2.16 Ventoline (2.5mg) Salbutamol/Sulphate
2.17 Casio Timer
3.0 Testing Items
3.1 Aerosol Particle Size Distribution Testing(By Malvern Spraytec)
3.2 Aerosol Particle Size Distribution Testing(By Marple 298 Cascade Impactor)
3.3 Nebulization Rate Testing(Including drugs testing)
3.4 Residual Volume Testing
3.5 Reliability Test
4.0 Testing Procedure
4.1 Aerosol Particle Size Distribution Testing(By Malvern Spraytec)
4.1.1 Each sample should be tested with 2.5ml 0.9% saline solution for 3 minutes.
4.1.2 Add 2.5ml 0.9% saline solution into the nebulizer kit,
4.1.3 Connect the nebulizer kit with NE100 and put at the testing site, the nebulizer
kit’s outlet must be kept at 3.0 cm from the laser beam.
4.1.4 Start recording Spraytec for more than 15 secs, then start NE100 for testing.
4.1.5 After 3.0 minutes have been reached, stop the NE100 and then stop Spraytec.
4.1.6 Checks Spraytec records
Fig 1. Malvern Spraytec and Testing site
2
4.2 Aerosol Particle Size Distribution Testing(By Marple 298 Cascade Impactor)
4.2.1 Each sample should be tested with 2.5 ml 2.5% NaF solution.
4.2.2 Add 2.5 ml 2.5% NaF solution into the nebulizer kit, measure the weight before
and after the testing.
4.2.3 Connect suction and sampling pumps to the cascade impactor testing module as
see in the Fig 2.
4.2.4 Connect the nebulizer kit with NE100 and connect the nebulizer kit outlet to
cascade impactor inlet.(Fig 2.)
4.2.5 The suction and sampling pumps are turned on and allowed to stabilize at
required flows.
4.2.6 Finally start the NE100.(Sampling times can be varied for different nebulizers to
allow for maximum deposit on each stage without coverloading stages.)
4.2.7 After sampling for the required time, NE100 is switched off, followed a few
seconds later by the sampling pump and then the suction pump.
4.2.8 Dismount the cascaade impactor from the testing module
4.2.9 Dismantle the impactor and determine the amount of NaF on the individual stages
of the impactor, the input connection and the outlet filter.
Fig 2. Cascade Impactor and Testing setup
4.3 Nebulization Rate Testing(Including Drug Testings)
4.3.1 Each sample should be tested with 2.0 ml 0.9% saline solution, Atrovent
Ipratropium Bromide, Atrovent Flixotide, AstraZeneca Terbutaline Sulphate and
Ventoline (2.5mg) Salbutamol/Sulphate for 1 minutes.
4.3.2 Add 2.0ml 0.9% saline solution/drugs into the nebulizer kit, measure the weight
before and after the testing.
4.3.3 Connect the nebulizer kit with NE100 and then start NE100 for testing.
4.3.4 After 1.0 minutes have been reached, stop the NE100
4.3.5 Calculates how many weight of the solution/drugs have been nebulizered
4.4 Residual Volume Testing
4.4.1 Each sample should be tested with 2.0ml 0.9% saline solution and nebulized till
the bottle is empty.
4.4.2 Add 2.0ml 0.9% saline solution into the nebulizer kit, measure the weight before
and after the testing.
4.4.3 Connect the nebulizer kit with NE100 and put at the testing site,
4.4.4 Start NE100
4.4.5 Shakes the nebulizer kit one or two times during nebulizing if there has large
droplet stick on the wall inside the nebulizer kit.
3
4.4.6 After the nebulizer kit is empty, stop NE100 and measure the mass of the tested
bottle
4.4.7 Calculates the Residual Volume
4.5 Reliability Test
4.5.1 Connect the tested NE100 with a nebulizer kit as loading.
4.5.2 Set a timer to control power on(30 minutes)/off(5 minutes) of tested NE100 to
repeat the working cycle as test specification request.
4.5.3 Before life test, and after every cumulative 100 working hours, check and record
the working pressure, working flow, current draw and the nebulization
performance.
5.0 Testing Results
5.1 Aerosol Particle Size Distribution Testing(By Malvern Spraytec)
NE100
Testing
times
Dv10 Dv50 Dv90 SMD SR
Sample 1
1 1.751 4.388 11.410 3.488
6.516
2 1.694 4.232 11.580 3.389
6.836
3 1.672 4.170 11.660 3.348
6.974
Fig 3
Mean
1.706 4.263 11.550 3.408 6.775
Std Dev
0.041 0.112 0.128 0.072 0.235
Sample 2
1 1.756 4.511 12.690 3.564
7.227
Fig 4
2 1.723 4.521 14.010 3.558
8.131
3 1.710 4.531 14.820 3.561
8.667
Mean
1.730 4.521 13.840 3.561 8.008
Std Dev
0.024 0.010 1.075 0.003 0.728


Specyfikacje produktu

Marka: Rossmax
Kategoria: inhalator
Model: NE100

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Rossmax NE100, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje inhalator Rossmax

Rossmax

Rossmax NB60 Instrukcja

12 Października 2024
Rossmax

Rossmax NL100 Instrukcja

7 Października 2024
Rossmax

Rossmax NE100 Instrukcja

4 Października 2024
Rossmax

Rossmax Car Charger Instrukcja

3 Października 2024

Instrukcje inhalator

Najnowsze instrukcje dla inhalator

Flaem

Flaem LightNeb Instrukcja

26 Grudnia 2024
Scala

Scala SC 9420 Instrukcja

5 Października 2024
Omron

Omron CompAIR NE-C801KD Instrukcja

1 Października 2024
Hydas

Hydas 3003.1.00 Instrukcja

23 Września 2024
Scala

Scala SC 350 Instrukcja

23 Września 2024
Microlife

Microlife NEB 10-N Instrukcja

23 Września 2024
Scala

Scala SC140 Instrukcja

20 Września 2024