Instrukcja obsługi Kyoritsu 2434


Przeczytaj poniżej 📖 instrukcję obsługi w języku polskim dla Kyoritsu 2434 (2 stron) w kategorii multimetr. Ta instrukcja była pomocna dla 4 osób i została oceniona przez 2 użytkowników na średnio 4.5 gwiazdek

Strona 1/2
INSTRUCTIONMANUAL
KEW SNAP
 
2434
KEW SNAP Series
LEAKAGECURRENTTESTER
KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD.,
TOKYO, JAPAN
1. SAFETY WARNINGS
This instrument has been designed and tested according to IEC
Publication 61010: Safety Requirements for Electronic Measuring
Apparatus. This instruction manual contains warnings and safety rules
which must be observed by the user to ensure safeoperation of the
instrumentandtoretainitinsafecondition.Therefore,readthroughthese
operatinginstructionsbeforestartingusingtheinstrument.
#
WARNING
Readthrough and understand instructions contained in this manual
beforestartingusingtheinstrument.
Saveandkeepthemanualathandtoenablequickreferencewhenever
necessary.
Besuretouse the instrumentonly in itsintended applicationsandto
followmeasurementproceduresdescribedinthemanual.
Be sure to understand and follow all safety instructions contained in
themanual.
Notfollowingtheaboveinstructionsmaycauseinjury,instrumentdamage
and/ordamagetoequipmentundertest.
Theinstrumentistobeusedonlyinitsintendedapplications.
Understandandfollowallthesafetyinstructionscontainedinthemanual.
Failuretofollowtheinstructionsmaycauseinjury,instrumentdamageand/
or damage toequipmentunder test. Kyoritsuisby nomeans liableforany
damageresultingfromtheinstrumentincontradictiontothiscautionarynote.
Thesymbol
#
indicatedontheinstrumentmeansthattheusermustrefer
torelatedpartsofthemanualforsafeoperationoftheinstrument.Besure
tocarefullyreadtheinstructionsfollowingeach
#
symbolinthismanual.
#
DANGER
is reservedforconditions andactions thatare likely to
causeseriousorfatalinjury.
#
WARNING
is reservedforconditions and actionsthatcancause
seriousorfatalInjury.
#
CAUTION
isreservedforconditionsandactionsthatcancause
minorinjuryorInstrumentdamage.
Following symbols are used on the instrument and in the instruction
manual.Attentionshouldbepaidtoeachsymboltoensureyoursafety.
#
Refertotheinstructionsinthemanual.
Indicatesaninstrumentwithdoubleorreinforcedinsulation.
Indicatesthatthisinstrumentcanclamponbareconductorswhen
measuringavoltagecorrespondingtotheapplicableMeasurement
category,whichismarkednexttothissymbol.
IndicatesAC(AlternatingCurrent).
IndicatesEarth.
#
DANGER
Never makemeasurementonacircuithavingpotential of 300VACor
greater.
Do not attempt to make measurement in the presence of flammable
gasses. Otherwise, the use of the instrument may cause sparking,
whichleadstoanexplosion.
The transformer jaws are made of metal and their tips are not
completelyinsulated.Beespeciallycarefulabout thepossibleshorting
wheretheequipmentundertesthasexposedmetalparts.
Neverattempttousetheinstrumentifitssurfaceoryourhandiswet.
Donotexceedthemaximumallowableinputofanymeasurementrange.
Neveropenthebatterycompartmentcoverwhenmakingmeasurement.
Never tryto make measurement if any abnormal conditions,such as
brokenTransformerjawsorcaseisnoted.
Theinstrumentistobeusedonlyinitsintendedapplicationsorconditions.
Otherwise,safetyfunctionsequippedwiththeinstrumentdoesn'twork,and
instrumentdamageorseriouspersonalinjurymaybecaused.
Keepyourfingersandhandsbehindthebarrierduringameasurement.
#
WARNING
Never attemptto make anymeasurement, if anyabnormal conditions
arenoted,suchasbrokencase,crackedtestleadsandexposedmetal
parts.
Do not install substitute parts or make any modification to the
instrument.  Return the instrument to Kyoritsu or your distributor for
repairorre-calibration.
#
CAUTION
Makesure that the range selector switch is set to an appropriate
positionbeforemakingmeasurement.
Donotexposethe instrumentto thedirect sun,extremetemperatures
ordewfall.
Be sure toset the range selector switch to the "OFF" position after
use.Whentheinstrumentwillnotbe in useforalong periodoftime,
placeitinstorageafterremovingthebatteries.
Usea dampcloth and detergent forcleaning the instrument. Donot
useabrasivesorsolvents.
Always switch off the instrument before opening the battery
compartmentcoverforbatteryreplacement.
Measurementcategories(Over-voltagecategories)
Toensuresafeoperationofmeasuringinstruments,IEC61010establishessafety
standards for various electricalenvironments, categorized as O toCAT IV, and
calledmeasurementcategories.
Higher-numbered categories correspond to electrical environmentswith greater
momentary energy,soameasuring instrumentdesigned for CATIIIenvironments
canenduregreatermomentaryenergythanonedesignedforCATII.
O: Circuitswhicharenotdirectlyconnectedtothemainspowersupply.
CATII: Primary electrical circuitsof equipmentconnectedto anACelectrical
outletbyapowercord.
CATIII: Primary electrical circuitsof the equipment connected directlyto the
distributionpanel,andfeedersfromthedistributionpaneltooutlets.
CATIV: The circuit fromthe service drop to the service entrance, and to the
power meter and primary overcurrent protection device (distribution
panel).
2. FEATURES
DigitalclamptesterforACleakagecurrentmeasurement.
Least affected by external magnetic field, providing wide measuring
rangefromverysmalltolargecurrents.
Designed to safety standardIEC 61010-2-032: measurement CAT.
,
300Vandpollutiondegree2.
Data hold function toallow for easy readings in dimly lit or hard-to-
reachlocations.
Providesfilteringfunctiontoremovehighfrequencygeneratedbysuch
equipmentasinverters.
Sleepfunctionpreventsunnecessarypowerconsumption.
Dynamicrangeof4000countsfullscale.
Largeeasy-to-readLCDdisplaywithletterheightof13mm.
Transformerjawsfittedwithguardtofurtherimprovesafety.
O: Device which is
not directly
connected to the
mains power supply
3. SPECIFICATIONS
Measuringrangesandaccuracy
Range MeasuringRange Accuracy(Frequencyrange)
400mA 0
399.9mA
 ±
2.0%rdg
±
4dgt
50/60Hz
 ±
3.0%rdg
±
5dgt
40
400Hz
4A 0
3.999A
100A 0
100.0A
Whenmeasuring current which pulse element is superposed,differences of the
indicatedvaluemaybecausedbetweenranges,ifthe peakvalue exceedsthe
measurement range to a large extent.In this case, the reading at the bigger
rangeshouldbetakenasarightvalue.
Sensitive transformer jawsare used forLeakage clampmeter.Becauseof the
characteristics of transformerjaws, which can beopened and closed, it is
impossibletoeliminate theinterferenceof externalmagneticfieldcompletely.If
there aresomething, which generating large magneticfield, at anearbysite,
current valuecan be displayed (
0
cannot be displayed.)before clampingon
theconductor.Forsuchacase,pleaseusetheinstrumentatalocationfarfrom
thething,whichgeneratingmagneticfield.
Followingarethetypicalthingsgeneratingmagneticfield.
Conductor fed large current, Motor, Equipment which has magnet, Integrating
wattmeter
Measuring method:
 DualIntegration
Display:
 Liquid crystal display with maximum reading of
3999
Low battery warning:
 "BATT"markappearsonthedisplay
Overrange Indication:
 "OL"appearsonthedisplaywhenupperlimitof
measuringrangeisexceeded
Response Time:
 Approx.2seconds
Sample Rate:
 Approx.2.5timespersecond
Location for use
Indooruse,Altitudeupto2000m
Accuracy-insured
 23
℃±
5
,relativehumidity85%orless
Temperature and
 (withoutcondensation)
Humidity Ranges:
Operating Temperature
 0-40
,relativehumidity85%orless
and Humidity Ranges:
 (withoutcondensation)
Storage Temperature
 -20-60
,relativehumidity85%orless
and Humidity Ranges:
 (withoutcondensation)
Power Source:
 Two1.5VR03(UM-4)batteries
Current Consumption:
 Approx.4mA
Measurement Time:
 Approx.150hours
Sleep Function:
Automatically powered down in about 10
minutesafterthelastswitchoperation
Safety Standard:
 IEC61010-1,IEC61010-2-032
measurement CAT.
300V, pollution degree 2
IEC61326(EMC),EN50581(RoHS)
Overload Protection:
 120Armsmax.for10seconds
Withstand Voltage:
3470Vrms (50/60Hz) for 5 seconds between
metalpartoftransformerjaws andhousing case
(excepttransformerjawcase)
Insulation Resistance:
 10M
Ω
or greaterat1000Vbetween metal part
of transformer jaws and housing case (except
transformerjawcase)
Conductor Size:
 Approx.28mmindiametermax.
Dimensions:
 169(L)
×
75(W)
×
40(D)mm
Weight:
  Approx.220gincludingbatteries
Accessories:
 TwoR03(UM-4)batteries
  CarryingcaseModel9097
  Instructionmanual
Optional Accessories:
 Multi-TranModel8008
4. INSTRUMENT LAYOUT
5. PREPARATIONS FOR MEASUREMENT
5-1 Checking Battery Voltage
Set the Range Selector Switch to any position other than the OFF
position. If the marks on the display is clearly visible without "BATT"
mark showing, batteryvoltageis OK.If thedisplayblanks or "BATT"is
indicated, replace the batteries according to section 8: Battery
Replacement.
NOTE
When the instrument is left powered on, the Sleep function
automaticallyshutthepower off;The display blankseven ifthe
RangeSelector Switch is set to a position other than the OFF
positioninthisstate.Topowerontheinstrument,turntheRange
SelectorSwitchorpressanybutton.Ifthedisplaystillblanks,the
batteriesarecompletelyexhausted.Replacethebatteries.
5-2 Checking Switch Setting
Make sure that the Range Selector Switch is set to the appropriate
range.
Alsomakesurethat dataholdfunctionisnotenabled. Ifinappropriate
rangeisselected,desiredmeasurementcannotbemade.
Tr an s fo rm J aw s
Tr ig ge r
Ra ng e S el ec to r Swi t ch
Ha nd Str a p
Di sp l ay
Frequency Selector Button
Data Hold Button
●LCD
Low Battery Warning
Frequency Response :
Wide
Frequency Response :
50/60Hz
Data Hold Indication
Unit
Ba rri e r
Barrier is a part providing protection against electrical shock and
ensuring the minimum required air and creepage distances.
3-phase 3-wire system
(In 4-wire system with neutral,
clamp onto all 4 wires)
Single-phase 2-wire system
(in 3-wire system with neutral,
clamp onto all 3 wires)
Load
Load
6. OPERATING INSTRUCTIONS
#
DANGER
In order to avoid possible shock hazard, never make measurement
oncircuitshavingapotentialof300VACorgreater.
The transformer jaws are made of metal and their tips are not
completely insulated. Be especially careful about the possible
shortingwheretheequipmentundertesthasexposedmetalparts.
Never make measurement with the battery compartment cover
removed.
Keepyourfingersandhandsbehindthebarrierduringmeasurement.
#
CAUTION
Take sufficient care to not to apply shock, vibration or excessive
force to the jaw tips.  Otherwise, precisely adjusted Transformer
Jawtipswillbedamaged.
When a foreign substance is stuck in the jaw tips or they cannot
properlyengage,thetransformerjawsdonotfullyclose.Insucha
case,donotreleasethejawtriggerabruptlyorattempttoclosethe
transformer jaws by applying external force.  Make sure that the
jaws close by themselves after removing the foreign substance or
makingthemfreetomove.
Themaximumsizeofaconductortobetestedis28mmindiameter.
Accuratemeasurementcannotbemadeonaconductorlargerthan
this,becausethetransformerjawscannotfullyclose.
Whenmeasuringlargecurrent,thetransformerjawsmaybuzz.This
hasnoeffectontheinstrument'sperformanceorsafety.
6-1 Leakage Current Measurement

(1)Set the Range Selector Switch to the desired position.  Current to
measureshouldbewithintheselectedmeasuringrange.
(2)Measuringoutofbalanceleakagecurrent(SeeFig.1):
Clampontoallconductorsexceptagroundedwire.Measuredcurrent
valueisshownonthedisplay.
(3)Measuringearthleakagecurrent(SeeFig.2):
Clamponto agroundedwire.Measured currentvalue isshown on the
display.
Fig.1Measuringoutofbalanceleakagecurrent
Fig.2Measuringearthleakagecurrent
6-2 How to Use Frequency Selector Button
Whenhigh frequencies from suchequipment as inverters are present in
the circuit under test, the instrument measures AC current of not only
50Hz or 60Hz of fundamental frequency but also of these high
frequenciesandharmonics.
To eliminate the effect of such high frequency noise and measure AC
currentof50Hzor60Hzfundamentalfrequency,a"high-cut"filtercircuit
in incorporated into the instrument which works when "50/60Hz"
frequencyresponseisselectedwiththeFrequencySelectorButton.Cut-
off frequency of the "high-cut" filter is about 160Hz with attenuation
characteristicofapprox.-24dB/octave.
When the Frequency Selector Button is pressed, "50/60Hz" mark is
shown on the left side of the display.  When the Frequency Selector
Button is pressed again, frequency response is switched to WIDE with
"WIDE"markshownonthedisplay.
OutputcharacteristicareshowninFig.3.
Fig.3KEWSNAP2434FrequencyCharacteristic
Note:
Characteristicof-24dB/octiavemeansthatsignalmagnitudedeclinesto
about one sixteenth of that at the initial frequency when frequency
doubles. KEW SNAP 2434 have the following two settings for the
FrequencySelectorButton.
WIDE(40Hz-):Permits measurement of currents of fundamental
frequencies as well as currents of high frequencies
generatedbysuchequipmentasinverters
50/60Hz(40-approx.160Hz):Filters out high frequency currents and
measurescurrentoffundamentalfrequency
only
100%
50%
20%
10%
5%
5k1k 2k500100 200502010
ー30
ー20
ー10
 0
Attenuation〔dB〕
 10
Frequency〔Hz〕
Percentage
Typical characteristic:
400mA/4A Range
Typical characteristic:
“WIDE”frequency response
Typical characteristic:
“50/60Hz”frequency response
high-cut filter -24dB/oct
Typical characteristic:
100A Range
Load
Grounded wire
Recently there has been increased use of power through inverters,
switching regulators, etc. When the high frequency noise from such
appliancesleaksorflowsintothegroundthroughcapacitorsnotfiltering
completely, the earth leakage breaker may trip even though there is no
"actual" leakage.  In such a case, the instrument do not give leakage
currentreadingif"50/60Hz"frequencyresponseisselected.
Takecurrentreadingswiththe50/60HzandWIDEfrequencyresponses
respectivelytomakeeffectiveuseoftheFrequencySelectorButton.
6-3LoadCurrentMeasurement
(1)
Set the Range Selector Switch to the desired position.
Current
to
measureshouldbewithintheselectedmeasuringrange.
(2)
Measuringloadcurrent(SeeFig.4):

Press the jaw trigger to open the transformer jaws and close
them
over one conductor only.Measured current value is shown
on the
display.
Fig.4Measuringloadcurrent
7. OTHER FUNCTIONS
7-1 Sleep Function
Thisisafunctiontopreventtheinstrumentfrombeingleftpowered
on
inordertoconservebatterylife.
The instrument automatically enters the Sleep (powered-down) Mode
about10minutesafterthelastswitchoperation.
To exit the Sle ep Mode, p ress th e Dat a Hold o r Freque ncy S elec tor
Button or turn th e Range Sel ector Switc h to O FF, then to any other
position.
HowtoExittheSleepMode:
TurntheRangeSelectorSwitchfrom
OFF
toanotherpositionwith
the Data Hold Butto n p ressed. Th en,
P.OFF
 is show n on the
display.ThisdisablestheSleepfunctionandenablescontinuoususeof
theinstrument.
To ena blet heS leep fu nctio n,t urn the Range Selector Switch ba ckto
OFF
,thentoanyotherposition.
Note:
Thein strum ent con sumess mallamou ntof cu rrent int heSleep M ode.
Whentheinstrumentisnotinuse,makesuretosettheRangeSelector
Switchto
OFF
.
7-2 Date Hold Function
Thisisafunctiontofreezethereadingsonthedisplay.WhentheData
Hol dBut ton is pre ssed onc e, th e current rea ding is held even thoug h
currentundertestvaries."H"markisshownontheupperrightcornerof
thedisplay.
Toexitthedataholdmode,presstheDataHoldButtonagain.
Note:WhentheSleepfunctionworkswhiletheinstrumentisinthedata
holdmode,dataholdiscancelled.
8. BATTERY REPLACEMENT
#
CAUTION
Donotmixnewandoldbatteries.
Install batteries in the orientation as shown inside the battery
compartment,observingcorrectpolarity.
In order to avoidpossible shock hazard, always set the Range
SelectorSwitchtotheOFFpositionbeforetryingtoreplacethe
batteries
When the battery voltage warning mark "BATT" is shown on the top
left corner of the LCD, replace the batteries.  Note that the display
blanks and"BATT" mark is not shown if the batteriesare completely
exhausted.
(1)SettheRangeSelectorSwitchto"OFF."
(2)Loosenthebattery-compartment-cover-fixingscrewonthelowerback
oftheinstrument.
(3)ReplacethebatterieswithtwonewR03(UM-4)1.5Vbatteries.
(4)Put the battery compartment
cover back in place and tighten
thescrew.
Note:Use alkaline batteries to perform
measuringforlong
9. OPTIONAL ACCESSORIES
Model8008(Multi-Tran)
These models help KEW SNAP 2434 to measure current greater than
1000Aortomakemeasurementonalargebus-barorconductor.
(1)SettheRangeSelectorSwitchto"100A".
(2)As shown, open the jaws and close them over the pickup coil of
Model8008.
(3)ClamponaconductorwithModel8008.
(4)Takethereadingandmultiplyitby10.
Note:Model 8008 cannot be usedfor leakage current measurement. For
detailed specifications, refer to the instruction manual for Model
8008.
Max.ConductorSize
MeasuringRange
CurrentTransformation
Ratio
M-8008
100mmindiameter
0
1000A 10:1
KEW2434
M- 8 0 0 8
Pick u p Coil
Conductor
Transformer
jaws
10
Screw Battery
Compartment
Cover
Batteries
92-1509B2-17
DISTRIBUTOR
Kyoritsu reserves the rightsto changespecifications ordesigns
describedinthismanualwithoutnoticeandwithoutobligations.
Load


Specyfikacje produktu

Marka: Kyoritsu
Kategoria: multimetr
Model: 2434

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Kyoritsu 2434, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje multimetr Kyoritsu

Kyoritsu

Kyoritsu KEW 2510 Instrukcja

30 Września 2024
Kyoritsu

Kyoritsu KEW 2500 Instrukcja

30 Września 2024
Kyoritsu

Kyoritsu KEW 2413R Instrukcja

30 Września 2024
Kyoritsu

Kyoritsu KT 200 Instrukcja

30 Września 2024
Kyoritsu

Kyoritsu 2431 Instrukcja

30 Września 2024
Kyoritsu

Kyoritsu 2434 Instrukcja

30 Września 2024
Kyoritsu

Kyoritsu KEW 2300R Instrukcja

29 Września 2024
Kyoritsu

Kyoritsu 1009 Instrukcja

27 Września 2024
Kyoritsu

Kyoritsu KEW2046R Instrukcja

20 Września 2024

Instrukcje multimetr

Najnowsze instrukcje dla multimetr

Voltcraft

Voltcraft PMM 3005-20 Instrukcja

9 Października 2024
Bearware

Bearware 306998 Instrukcja

9 Października 2024
Bearware

Bearware 307001 Instrukcja

9 Października 2024
Fluke

Fluke 283 FC Instrukcja

9 Października 2024
Noyafa

Noyafa NF-5330C Instrukcja

9 Października 2024
Noyafa

Noyafa NF-5310B Instrukcja

9 Października 2024
Emos

Emos M3216 Instrukcja

8 Października 2024
Uni-T

Uni-T UT121B Instrukcja

7 Października 2024
Maxwell

Maxwell MX-25103 Instrukcja

6 Października 2024
Velleman

Velleman DVM840 Instrukcja

5 Października 2024