Instrukcja obsługi Testo 6381


Przeczytaj poniżej 📖 instrukcję obsługi w języku polskim dla Testo 6381 (4 stron) w kategorii Niesklasyfikowane. Ta instrukcja była pomocna dla 3 osób i została oceniona przez 2 użytkowników na średnio 4.5 gwiazdek

Strona 1/4
Measurement of differential pressure, flow velocity, volume
flow; optional: humidity and temperature
Automatic zero-point adjustment guarantees high,
temperature-independent accuracy and long-term stability
Low measurement range up to 10 Pa ensures very high
precision at lowest pressures
Ethernet, relay and analog outputs allow optimum
integration into individual automation systems
The P2A software for parameterization, adjustment and
analysis saves time and costs in commissioning and
maintenance
Configurable alarm management with adjustable response
delay and alarm acknowledgement
Dierential pressure transmitter
with humidity/temperature option
testo 6381
The differential pressure transmitter testo 6381 was
developed specially for monitoring differential pressure in
the measuring range from 10 Pa to 1000 hPa. In cleanroom
technology, the maintenance of positive pressure prevents
the entry of contaminated air. In addition to this, the flow
velocity or the volume flow can be calculated from the
measurement of the differential pressure in a Pitot tube.
Thanks to an optional probe from the probe series 6610, the
additional recording of humidity and temperature with one
instrument is also possible.
The testo 6381 is particularly outstanding thanks to the
automatioc zero-point adjustment which ensures high
accuracy and long-term stability.
The integrated self-monitoring and early warning function
also guarantees the operator high system availability.
www.testo.com
hPa
%RH
°C
testo 6381 Dierential pressure transmitter with
humidity/temperature option
Technical data
Supply
Analog outputs
Other outputs
Meas. cycle 1/sec
Quantity Standard: 1;
with optional humidity probe: 3
Scaling Differential pressure: scalable ±50% of
measuring range final value;
freely scalable within measuring range
Resolution 12 bit
Max. load max. 500 W
Humidity/temperature optional
Parameters
Probe
Type
Parameters
Measurement uncertainty*
%RH / °C/°F / °C
td / °Ftd / g/kg / gr/lb / g/m3 / gr/ft³ / ppmV
/ °Cwb / °Fwb / kJ/kg / mbar / inch H2O / °Ctm (H2O2)/°Ftm
(H 2O2) / % Vol
for deviations from media temp. ±25 °C: ±0.02 %RH/K
Meas. range
Humidity /
trace humidity
Humidity
Dewpoint
Temperature
Temp. at
+25°C /
+77°F
Inputs/outputs
Output type 0/4 to 20 mA (4-wire) (24 VAC/DC)
0 to 1/5/10 V (4-wire) (24 VAC/DC)
Ethernet Optional
Relay Optional: 4 relays (free allocation to
measurement channels or as collective alarm
in operating menu/P2A), up to 250 VAC/3A
(NO or NC)
Digital Mini-DIN for P2A software
Voltage supply 20 to 30 VAC/DC, 300 mA current
consumption, galvanically separate signal and
supply line
Wall
testo 6612testo 6611 testo 6613 testo 6614 testo 6615 testo 6617
Channel Channel Duct
heated
Cable
trace
humidity
Cable
with cover
electrode
monitoring
testo 6611 testo 6612 testo 6613 testo 6614 testo 6615 testo 6617
±(1.0 + 0.007 * mv) %RH for 0 to 100
%RH / ±(1.4 + 0.007 * mv) %RH for
90 to 100 %RH
±(1.0+ 0.007
* mv) %RH
for 0 to 100
%RH
±(1.2 + 0.007 * mv)
%RH for 0 to 90
%RH / ±(1.6 + 0007
* mv) %RH for 90 to
100 %RH
±0.15 °C/
32.2 °F
Pt1000 Class AA
±0.15 °C/
32.2 °F
Pt100 Class AA
±0.15 °C/
32.2 °F
Pt1000
Class AA
0 to 100 %RH -60 to +30
°C td
0 to 100
%RH
-20 to +70 °C
-4 to +158 °F
-30 to +150 °C
-22 to +302 °F
-40 to +180 °C
-40 to +356 °F
-40 to +120 °C
-40 to +248 °F
-40 to +180 °C
-40 to +356 °F
Dierential pressure
Parameters
Measuring range 0 to 10 Pa
0 to 50 Pa
0 to 100 Pa
0 to 500 Pa
0 to 10 hPa
0 to 50 hPa
0 to 100 hPa
0 to 500 hPa
0 to 1000 hPa
-10 to 10 Pa
-50 to 50 Pa
-100 to 100 Pa
-500 to 500 Pa
-10 to 10 hPa
-50 to 50 hPa
-100 to 100 hPa
-500 to 500 hPa
-1000 to 1000 hPa
Measurement
uncertainty*
±0.5% of measurement range final
value
±0.3 Pa
Temperature gain drift: 0.03% of
measuring range per Kelvin deviaton
from nominal temperature 22 °C
Zero-point: 0% (thanks to cyclic
zero-point adjustment)
Selectable units Differential pressure in Pa, hPa, kPa,
mbar, bar, mmH2O, kg/cm2, PSI, inch
HG, inch H2O
calculated parameters: volume flow
in m3/h, l/min, Nm3/h, Nl/min
Flow velocity in m/s, ft/min
Sensor Piezoresistive sensor
Autom. zero-point
adjustment
via magnetic valve
Frequency adjustable: 15 sec, 30
sec, 1 min, 5 min, 10 min
Overload
* The determination of measurement uncertainty takes
place according to GUM (Guide to the Expression of
Uncertainty in Measurement):
For the determination of measurement uncertainty, the
accuracy of the measuring instrument (hysteresis, linearity,
reproduceability), the uncertainty contribution of the test site as
well as the uncertainty of the adjustment site (works calibration)
are taken into account. For this purpose, the value of k=2 of the
extension factor, which is usual in measurement technology is
used as a basis, which corresponds to a trust level of 95%.
Measuring range Overload
0 to 10 Pa 20000 Pa
0 to 50 Pa 20000 Pa
0 to 100 Pa 20000 Pa
0 to 500 Pa 20000 Pa
0 to 10 hPa 200 hPa
0 to 50 hPa 750 hPa
0 to 100 hPa 750 hPa
0 to 500 hPa 2500 hPa
0 to 1000 hPa 2500 hPa
-10 to 10 Pa 20000 Pa
-50 to 50 Pa 20000 Pa
-100 to 100 Pa 20000 Pa
-500 to 500 Pa 20000 Pa
-10 to 10 hPa 200 hPa
-50 to 50 hPa 750 hPa
-100 to 100 hPa 750 hPa
-500 to 500 hPa 2500 hPa
-1000 to 1000 hPa 2500 hPa
±1 K at 0
°C td
±2 K at -40
°C td
±4 K at -50
°C td
+-+-+-+-
+
-
CH1 CH2 CH324V
1
20-30 V
AC/DC
0V
testo 6381 Dierential pressure transmitter with
humidity/temperature option
Technical data / Technical drawings / Connection
plan
Technical drawings
Connection plan
Display
Model
Miscellaneous
Display optional: 3-line LCD with
multi-language operating menu
0.1 %RH
0.01 °C / 0.01 °F
General technical data
Material Metal housing
Dimensions 162 x 122 x 77 mm
Resolution
Differential pressure
Humidity
Temperature
Measuring range Resolution
Operation temperatureWith / without
display
-5 to 50 °C / 23 to 122 °F
Storage temperature -20 to 60 °C / -4 to 140 °F
Process temperature -20 to +65 °C / -4 to +149 °F
Protection class IP 65
EMC EU guideline 2004/108/EC
Operating conditions
Weight 1.96 kg; optional: Ethernet
intermediary layer 0.61 kg
0 to 10 Pa 0.1 Pa
0 to 50 Pa 0.1 Pa
0 to 100 Pa 0.1 Pa
0 to 500 Pa 0.1 Pa
0 to 10 hPa 0.01 hPa
0 to 50 hPa 0.01 hPa
0 to 100 hPa 0.1 hPa
0 to 500 hPa 0.1 hPa
0 to 1000 hPa 1 hPa
-10 to 10 Pa 0.1 Pa
-50 to 50 Pa 0.1 Pa
-100 to 100 Pa 0.1 Pa
-500 to 500 Pa 0.1 Pa
-10 to 10 hPa 0.01 hPa
-50 to 50 hPa 0.01 hPa
-100 to 100 hPa 0.1 hPa
-500 to 500 hPa 0.1 hPa
-1000 to 1000 hPa 1 hPa
Connection nipple Ø 6 mm --> suitable hoses 4
mm + 4.8 mm


Specyfikacje produktu

Marka: Testo
Kategoria: Niesklasyfikowane
Model: 6381

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Testo 6381, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Niesklasyfikowane Testo

Testo

Testo 6681 Instrukcja

27 Września 2024
Testo

Testo 6651 Instrukcja

27 Września 2024
Testo

Testo 550 Instrukcja

27 Września 2024
Testo

Testo 6621 Instrukcja

27 Września 2024
Testo

Testo 6631 Instrukcja

27 Września 2024
Testo

Testo 557 Instrukcja

27 Września 2024
Testo

Testo 184 H1 Instrukcja

26 Września 2024
Testo

Testo 6381 Instrukcja

22 Września 2024
Testo

Testo 184 Instrukcja

21 Września 2024
Testo

Testo 6321 Instrukcja

21 Września 2024

Instrukcje Niesklasyfikowane

Najnowsze instrukcje dla Niesklasyfikowane

Stihl

Stihl HT 70 Instrukcja

15 Października 2024
Ernesto

Ernesto H14270 Instrukcja

15 Października 2024
Uniden

Uniden BT MIC KIT Instrukcja

15 Października 2024
Vaude

Vaude Omnis Bike 26 Instrukcja

15 Października 2024
ART

ART ProChannel II Instrukcja

15 Października 2024
Aukey

Aukey DR02J Instrukcja

15 Października 2024
Master Lock

Master Lock 653EURD Instrukcja

15 Października 2024
Roland

Roland FR-18 Diatonic Instrukcja

15 Października 2024