Digital Automatic Blood Pressure Monitor
Model MIT Elite
• Instruction Manual
• Mode d’emploi
• Gebrauchsanweisung
• Manuale di istruzioni
• Manual de instrucciones
• Gebruiksaanwijzing
•
IM-HEM-7300-WE7-03-05/2012
5329377-2D
12D0973
EN
FR
DE
IT
ES
NL
RU
AR

2
Contents
Before using the unit
Introduction ................................................................................ 3
Important Safety Information .................................................... 4
1. Overview ............................................................................... 6
2. Preparation ......................................................................... 10
2.1 Installing/Replacing the Batteries.................................... 10
2.2 Setting the Date and Time .............................................. 12
Operating instructions
3. Using the Unit..................................................................... 15
3.1 How to Sit Correctly When Taking a Measurement ........ 15
3.2 Applying the Arm Cuff ..................................................... 17
3.3 Taking a Reading............................................................ 20
3.4 Using the Memory Function ............................................25
4. Quick Reference Guide ..................................................... 29
Care and maintenance
5. Handling Errors and Problems ......................................... 30
5.1 Error Messages...............................................................30
5.2 Troubleshooting ..............................................................32
6. Maintenance and Storage ................................................. 34
7. Optional Parts .................................................................... 36
8. Technical Data.................................................................... 37
9. Some Useful Information about Blood Pressure ............ 39

3
EN
Introduction
Thank you for purchasing the OMRON MIT Elite Upper Arm Blood
Pressure Monitor.
The OMRON MIT Elite is a fully automatic blood pressure monitor,
operating on the oscillometric principle. It measures your blood
pressure and pulse rate simply and quickly. For comfortable controlled
inflation without the need for pressure pre-setting or re-inflation the
device uses its advanced “IntelliSense” technology.
The unit also stores up to 90 measurements in memory and calculates
an average reading based on the three most recent measurements
taken within 10 minutes of the last reading.
Please read this instruction manual thoroughly before
using the unit. For specific information about your own
blood pressure, CONSULT YOUR DOCTOR.
g

4
Important Safety Information
Important Safety Information
Consult your doctor during pregnancy, arrhythmia and arteriosclerosis. Please
read this section carefully before using the unit.
Warning:
Indicates a potentially hazardous situation which, if not avoided, could result in
death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and self-
treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
doctor before using the unit. Cuff inflation can cause internal bleeding.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
(Optional AC Adapter Usage)
• Never plug in or unplug the power cord from the electric outlet with wet
hands.
Caution:
Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment or
other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone, or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
• Do not operate unit in a moving vehicle (car, airplane).
(Optional AC Adapter Usage)
• Use only the original AC adapter designed for this unit. Use of unsupported
adapters may damage and/or may be hazardous to the unit.
• Plug the AC adapter into the appropriate voltage outlet.

6
1.Overview
1. Overview
Main unit
A. Air jack
B. Date/time setting button ( )
C. O/I START button
D. Memory button ( )
E. AC adapter jack (for optional AC
adapter)
F. Display
G. Battery compartment
F
C
A
G
E
D
B

7
1.Overview
EN
Arm Cuff
H. Arm cuff
(Medium cuff: arm circumference 22-32 cm)
I. Air tube
J. Air plug
J
I
H

8
1.Overview
Display
K. Systolic blood pressure
L. Diastolic blood pressure
M. Heartbeat symbol
1.
Flashes during measurement
2.
If flashing after measurement
complete indicates blood
pressure out of recommended
range
N. Deflation symbol
O. Memory symbol
Displayed when viewing values
stored in memory
P. Average value symbol
Displayed when viewing value for
the last three measurements
Q. Battery low symbol
R. Movement error symbol
Displayed if you move your body
during the measurement
S. Pulse display
T. Date/Time display
U. Irregular heartbeat symbol
V. Guest symbol
K
L
N
P
R
M
T
V
S
U
O
Q

9
1.Overview
EN
Package contents
W. Storage case
X. Four “AAA” alkaline (LR03)
batteries
• Instruction manual
• Guarantee card
• Blood pressure pass
XW

10
2.Preparation
2. Preparation
2.1 Installing/Replacing the Batteries
1. Turn the main unit upside down.
2. Slide the battery cover in the
direction of the arrow while
pressing the ribbed part of the
cover.
3. Install or replace four “ ”AAA size
batteries so that the + (positive)
and - (negative) polarities match
the polarities indicated on the
battery compartment.
4. Put the battery cover back in
place.
Slide the battery cover as
indicated, until it clicks into
place.
Note: The measurement values continue to be stored in
memory even after the batteries are replaced.

11
2.Preparation
EN
Battery Life & Replacement
If the battery low symbol ( ) appears on the display, replace all four
batteries at the same time.
- When the battery low symbol ( ) starts to blink, you will still be
able to use the unit for a short while. You should replace the
batteries with new ones ahead of time.
- When the symbol ( ) remains lit, the batteries are exhausted. You
should replace the batteries with new ones at once. Turn the unit off
before replacing the batteries.
• Remove the batteries if the unit will not be used for three months or
more.
• If the batteries have been removed for more than 30 seconds, the
Date/Time setting will need to be reset. See “2.2 Setting the Date and
Time” for details.
• Dispose of batteries according to applicable local regulations.
Four new “ ”AAA alkaline batteries will last for approximately 300
measurements, when used to take three measurements a day.
Since the supplied batteries are for monitoring use only, they may have
a shorter life and not last for 300 measurements.

12
2.Preparation
2.2 Setting the Date and Time
Your blood pressure monitor automatically stores up to 90
measurements values with the date and time. To make use of the
memory and average values functions:
• Set the unit to the correct date and time before taking a measurement
for the first time.
• If the batteries have been removed for more than 30 seconds, the
Date/Time setting will need to be set.
1. Press and hold the Date/time setting
button ( ) to adjust the date and
time. The year digits (2008) will flash
on the display.
2. Press the Memory button ( ) to
advance the digits one at a time.
Notes:
• The range for the year setting is
2008 to 2030. If the year reaches
2030, it will return to 2008.
• If you hold down the Memory
button ( ), the digits will advance
rapidly.

14
2.Preparation
6. Repeat steps 2 and 3 to set the
hour.
The hour is set and the minutes
digits flash on the display.
7. Repeat steps 2 and 3 to set the minutes.
The minutes settings is set.
Press the Date/time setting button
( ) to turn the monitor off.
When adjusting the date and time,
press the Date/time setting button
( ) to turn the monitor on, follow
the steps above, and press the O/I START button to turn the
monitor off.

15
3.Using the Unit
EN
3. Using the Unit
3.1
How to Sit Correctly When Taking a Measurement
You can take a measurement on either your left or right arm.
Notes:
• Measurements should be taken in a quiet place and you should be in
a relaxed, seated position. Make sure that the room is not too hot or
cold.
• Avoid eating, drinking alcohol, smoking, or exercising for at least 30
minutes before taking a measurement.
• Do not move or talk during measurement.
Correct
Posture
Remove tight fitting clothing from
your upper arm, and any thick
clothing such as a sweater.
Do not place the cuff over thick
clothes and do not roll up your sleeve
if it is too tight.
Sit upright
with your
back straight.
Place your arm on a table so
that the cuff will be at the same
level as your heart.
The gap between the
chair and the top of the
table should be between
25cm to 30cm.
p g

16
3.Using the Unit
Notes:
• Correct posture during measurement is necessary to get accurate
results.
• You should also try to measure your blood pressure at the same time
each day. (Within 1 hour after waking up is recommended.)
Incorrect Posture
These situations could lead to higher blood pressure values due to
strain or the arm cuff being lower than the heart.
If the arm cuff is at a lower position than your heart use cushions
etc., to adjust the height of your arm.
•Arched back (leaning
forwards)
•Sitting cross-legged
•Sitting on a sofa or at a low
table so that you tend to lean
forward

17
3.Using the Unit
EN
3.2 Applying the Arm Cuff
Notes:
• Be sure to wrap the arm cuff correctly so that you get accurate
results.
• Measurements can be taken in light clothing. However, please
remove thick clothes, such as sweaters, before taking a reading.
• You can take a measurement on either your left or right arm. The
blood pressure can differ between the right arm and the left arm and
therefore also the measured blood pressure values can be different.
Omron recommends to always use the same arm for measurement. If
the values between the two arms differ substantially, please check
with your doctor which arm to use for your measurement.
Taking measurements on the left arm
1. Insert the air plug into the air
jack on the left side of the main
unit.

18
3.Using the Unit
2. Put your left arm through the cuff
loop.
Note: If the cuff is not assembled,
pass the end of the cuff
furthest from the tubing
through the metal D-ring to
form a loop. The smooth cloth
should be on the inside of the
cuff loop.

19
3.Using the Unit
EN
3. Position the arm correctly.
Taking measurements on the right arm
Apply the cuff so that the
air tube is at the side of
your elbow.
• Be careful not to rest
your arm on the air tube,
or otherwise restrict the
flow of air to the cuff.
• The cuff should be 1 to
2 cm above the elbow.
4. When the cuff is positioned correctly, close the fabric fastener
FIRMLY.
1) The air tube should run
down the inside of your
forearm and be in line with
your middle finger.
2) The bottom of the cuff
should be approximately 1
to 2 cm above your elbow.
3) Apply the cuff to your
upper arm so that the
coloured marker (blue
arrow under tube) is
centred on the middle of
your inner arm and points
down the inside of the
arm.
1 to 2 cm

20
3.Using the Unit
3.3 Taking a Reading
1. Press the O/I START button to turn on the unit.
All symbols appear on the display.
The cuff starts to inflate automatically. As the cuff inflates, the
unit automatically determines your ideal inflation level. This unit
detects the pulse during inflation. Do not move your arm and
remain still until the entire measurement process is completed.
Note: To stop the inflation or measurement, press and release
the O/I START button. The unit will stop inflating, start
deflation and will turn off.
2. When the measurement is complete, the arm
cuff completely deflates. Your blood pressure
and pulse rate are displayed.

22
3.Using the Unit
Using the Guest Mode
The monitor stores measurement values for one user in the memory.
The guest mode can be used to take a single measurement for another
user.
No measurement values are stored in the memory when the guest
mode is selected.
1. Press and hold the O/I START button for 3 seconds.
The Guest symbol ( ) appears on
the display.
2. Release the O/I START button.
The cuff starts to inflate automatically.
Note: To stop the inflation, press the O/I START button. The
monitor will stop inflating, start deflating and turn off.
3. Inflation stops and the measurement is started. When the
measurement is completed, your blood pressure and pulse rate
appear on the display.

23
3.Using the Unit
EN
Important:
• If your systolic or diastolic pressure is outside the
standard range, the heartbeat symbol will blink
when the measurement result is displayed.
Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
This criteria is for home blood pressure measurement.
For professional office blood pressure measurement criteria, please
refer to Chapter 9 “Some Useful Information about Blood Pressure”.
• Your blood pressure monitor includes an irregular
heartbeat feature. Irregular heartbeats can
influence the results of the measurement. The
irregular heartbeat algorithm automatically
determines if the measurement is usable or
needs to be repeated. If the measurement results
are affected by irregular heartbeats but the result
is valid, the result is shown together with the
irregular heartbeat symbol. If the irregular
heartbeats cause the measurement to be invalid, no result is shown.
If the irregular heartbeat symbol ( ) is shown after you have taken
a measurement, repeat the measurement. If the irregular heartbeat
symbol is shown frequently, please make your doctor aware of it.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg

24
3.Using the Unit
What is Irregular Heartbeat?
An irregular heartbeat is a
heartbeat rhythm that varies by
more than 25% from the average
heartbeat rhythm detected while
the unit is measuring the systolic
and diastolic blood pressure.
If such an irregular rhythm is
detected more than twice during
measurement, the irregular
heartbeat symbol ( ) appears on the symbol when the
measurement results are displayed.
What is Arrhythmia?
A heartbeat is stimulated by electrical signals that cause the heart
to contract.
Arrhythmia is a condition where the heartbeat rhythm is abnormal
due to flaws in the bio-electrical system that drives the heartbeat.
Typical symptoms are skipped heartbeats, premature contraction,
an abnormally rapid (tachycardia) or slow (bradycardia) pulse. This
can be caused by heart disease, aging, physical predisposition,
stress, lack of sleep, fatigue etc. Arrhythmia can only be diagnosed
by a doctor through a special examination.
Whether the appearance of the irregular heartbeat symbol ( ) in
the results indicates arrhythmia or not can only be determined by
an examination and diagnosis by your doctor.
Warning:
If the Irregular heartbeat symbol ( ) is shown frequently, please
make your doctor aware of it. Conducting self-diagnosis and
treatment based on measurement results are dangerous. Be sure to
follow the instructions of your doctor.
Normal Heartbeat
Irregular Heartbeat
Pulse
Blood pressure
Blood pressure
Short Long
Pulse

25
3.Using the Unit
EN
3.4 Using the Memory Function
This unit has a memory capable of storing 90 sets of readings. Every
time you complete the measurement, the unit automatically stores the
blood pressure and pulse rate.
If there are three readings within 10 minutes of the last reading, the
average of the three readings is displayed.
Note: When 90 sets of readings are stored in memory, the oldest set
will be deleted to store a new set.
1. Press the Memory button ( ).
If there are three readings stored in memory that were taken
within 10 minutes of the last reading, an average value for
those reading is displayed. (If there are only two readings in
memory for that period, the average will be based on two
readings. If there is only one reading in memory for that period,
that reading will be displayed as the average value.)
Note: If there are no measurements
results stored in memory, the screen
to the right is displayed.

26
3.Using the Unit
2. While the average value is
displayed, press the Memory button
( ) again.
The result from the most recent
measurement is displayed.
The date and time are
alternately displayed.
Note: If your systolic or diastolic
pressure is outside the standard
range, the heartbeat symbol will
blink when the measurement result
is displayed. Refer to section 3.3.

27
3.Using the Unit
EN
3. Press the Memory button ( )
repeatedly to cycle through the
previous measurement results.
Keep the button pressed down to
cycle rapidly through the previous
results.
If the irregular heartbeat symbol was
displayed at the time of a
measurement, this is displayed when that result is displayed.
4. Press the O/I START button to turn the unit off.
If you forget to turn the unit off, it will automatically shut itself off
after two minutes.

28
3.Using the Unit
To Delete All the Values Stored in Memory
You cannot delete individual stored readings, all the readings in the
unit will be deleted.
1. Press the Memory button
( ) to turn the monitor on.
2. Press the Memory button
( ) again. While holding the
Memory button ( ), press
the O/I Start button down for
approximately 5 seconds
until the digital display
appears as it does in the illustration to the right.
3. Press the O/I START button to turn the unit off.
If you forget to turn the unit off, it will automatically shut itself off
after two minutes.
Note: If the date and time are reset to a time before the most
recent measurement, the average value will be based on
any measurements taken after the date and time were
reset. However, you can still view the readings stored in
memory.

29
4.Quick Reference Guide
EN
4. Quick Reference Guide
If you are using this device for the first time, please read carefully
chapter 3 of this instruction manual. To help ensure a reliable reading,
avoid eating, drinking alcohol, smoking, or exercising for at least 30
minutes before taking a measurement.
Note: Remove any tight-fitting clothing from your upper arm.
1.
Sit on a chair with your feet flat on the floor and place your arm on
a table so that the arm cuff will be at the same level as your heart.
2. Apply the arm cuff to your upper arm. The coloured marker
should be centred on the inside of your arm and point down the
inside of the arm, so that the air tube runs down the inside of
your forearm and is in line with your middle finger.
3. Secure the cuff around your arm using the fabric fastener strip.
4. Press the O/I START button.
When measurement is complete,
the monitor displays your blood
pressure and pulse rate, and
automatically deflates the cuff.
Notes:
• Always wait at least 2-3 minutes before taking another blood
pressure measurement.
• Please note that all measurements are stored in the memory.
If different people measure with the same device make sure
that you are aware of this fact.

30
5.Handling Errors and Problems
5. Handling Errors and Problems
5.1 Error Messages
Error Display Cause Remedy
Movement during
measurement
Repeat measurement. Remain still
and do not talk during
measurement.
Refer to section 3.3.
Air plug disconnected. Insert the air plug securely.
Refer to section 3.2.
Arm cuff not applied
correctly.
Apply the arm cuff correctly.
Refer to section 3.2.
Clothing is interfering with
the arm cuff.
Remove any clothing interfering
with the arm cuff.
Refer to section 3.2.
Air is leaking from the arm
cuff.
Replace cuff with new one.
Refer to Chapter 7.
The arm cuff was inflated
above 299 mmHg when
inflating the cuff manually.
Remove the arm cuff and take
another measurement.
Refer to section 3.3.
Battery power is low.
Replace all four “AAA” batteries
with new ones.
Refer to section 2.1.
Blinks or appears
continuously

31
5.Handling Errors and Problems
EN
Note: The irregular heartbeat symbol may also be displayed with error messages.
Device error.
Contact your OMRON retail outlet
or distributor.
Refer to Chapter 8.
Error Display Cause Remedy

32
5.Handling Errors and Problems
5.2 Troubleshooting
Problem Cause Remedy
The reading is
extremely low (or high).
Arm cuff not applied
correctly.
Apply the arm cuff
correctly.
Refer to section 3.2.
Movement or talking
during measurement.
Remain still and do not
talk during
measurement.
Refer to section 3.3.
Clothing is interfering
with the arm cuff.
Remove any clothing
interfering with the arm
cuff.
Refer to section 3.1.
Arm cuff pressure does
not rise.
Is the air jack securely
inserted into the main
unit?
Make sure that the air
jack is connected
securely.
Refer to section 3.2.
Air is leaking from the
arm cuff.
Replace the arm cuff
with a new one.
Refer to chapter 7.
Arm cuff deflates too
soon. The arm cuff is loose.
Apply the cuff correctly
so that it is firmly
wrapped around the
arm.
Refer to section 3.2.
Cannot measure or
readings are too high.
Is the cuff wrapped
correctly around the
arm?
Attach the arm cuff
correctly.
The cuff pressure does
not rise although the
pump motor can be
heard.
Check that the air plug
is properly connected to
the unit. Push the air
plug firmly into the air
jack.
The unit loses power
during measurement. The batteries are empty. Replace the batteries
with new ones.

33
5.Handling Errors and Problems
EN
Nothing happens when
you press the buttons.
The batteries are empty. Replace the batteries
with new ones.
The batteries have been
inserted incorrectly.
Insert the batteries with
the correct (+/ -) polarity.
Other problems.
Press the O/I START button and repeat
measurement.
If the problem continues, try replacing the batteries
with new ones.
If this still does not solve the problem, contact your
OMRON retail outlet or distributor.
Problem Cause Remedy

34
6.Maintenance and Storage
6. Maintenance and Storage
Maintenance
To protect your unit from damage, please observe the following:
• Do not subject the main unit and cuff to extreme temperatures,
humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
• Do not inflate the arm cuff over 299 mmHg.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations (for example,
dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit.
• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean the arm cuff.
• Do not carry out repairs of any kind by yourself. If a defect occurs,
consult your OMRON retail outlet or distributor as mentioned on the
packaging.
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean the arm cuff.
Calibration and Service
• The accuracy of this blood pressure monitor has been carefully tested
and is designed for a long service life.
• It is generally recommended to have the unit inspected every two
years to ensure correct functioning and accuracy. Please consult your
authorised OMRON dealer or the OMRON Customer Service at the
address given on the packaging or attached literature.
Note: Inspections are normally not covered by guarantee, please
check the included guarantee card.

35
6.Maintenance and Storage
EN
• Do not carry out any repairs yourself. If a defect occurs or you have
doubts about the correct functioning of the device, consult your
authorised OMRON dealer or the OMRON Customer Service.
Storage
Keep the unit in its storage case when not in use.
1. Unplug the air tube from the air jack.
2. Gently fold the air tube into the arm
cuff.
Note: Do not bend the air tube
excessively.
3. Place the arm cuff and main unit in
the storage case.
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct sunlight,
dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an
angle.

36
7.Optional Parts
7. Optional Parts
Using the Optional AC Adapter
Insert batteries into the battery compartment even when using the AC
adapter.
Notes:
• Never unplug the power cord with wet hands.
• Use only the original AC adapter designed for this unit. Use of
unsupported adapters may damage the unit.
• When storing the optional AC adapter with the main unit, be
careful not to damage the main unit or cuff.
1. Insert the AC adapter plug
into the AC adapter jack on
the right side of the main unit.
2. Plug the AC adapter into an
electrical outlet.
To disconnect the AC adapter, unplug the AC adapter from the
electrical outlet first and then immediately remove the AC adapter plug
from the main unit.
Medium Arm Cuff
Arm circumference 22 - 32 cm
Large Arm Cuff
Arm circumference 32 - 42 cm
AC Adapter
CM1-9997578-9
CL-MIT Elite
9999358-2
Note: CL1 cuff cannot be
used for this device.
Adapter-3094298-6
(Model: AC Adapter-E1600)
Note: Q Adapter-1098336-8 and
R Adapter-9997605-0
cannot be used for this
device.
1
2

37
8.Technical Data
EN
8. Technical Data
Note: Subject to technical modification without prior notice.
• This device fulfils the previsions of EC directive 93/42/EEC (Medical Device Directive).
• This blood pressure monitor is designed according to the European Standard EN1060,
Non-invasive sphygmomanometers Part 1: General Requirements and Part 3:
Supplementary requirements for electromechanical blood pressure measuring
systems.
• This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co. Ltd., Japan. The Core component for OMRON blood pressure
monitors, which is the Pressure Sensor, is produced in Japan.
Product Description Digital Automatic Blood Pressure Monitor
Model OMRON MIT Elite (HEM-7300-WE7)
Display LCD Digital Display
Measurement Method Oscillometric method
Measurement Range Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180/min.
Accuracy Pressure: r3 mmHg
Pulse: r5% of display reading
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic pressure release valve
Memory 90 Measurements with date and time
Power Source 4 “AAA” alkaline batteries 1.5V or AC/ DC adapter
(optional, 6V 4W)
Battery life Approximately 300 uses when used 3 times a day with new
alkaline batteries when inflated to 170mmHg at 23°C
Operating temperature/
Humidity
+10°C to +40°C
Maximum: 30 to 85% RH
Storage temperature/
Humidity/Air Pressure
-20°C to +60°C
Maximum: 10 to 95% RH
700 - 1060hPa
Console Weight Approximately 240g without batteries
Cuff Weight Approximately 130g
Outer Dimensions Approximately 74 (w) mm u u 30.6 (h) mm 157 (l) mm
Cuff Dimensions Approximately 146 mm u 446 mm
(Medium cuff: arm circumference 22 to 32 cm)
Package Content Medium cuff, instruction manual, storage case, battery set,
guarantee card, blood pressure pass
= Type B
= Type B
Specyfikacje produktu
Marka: | Omron |
Kategoria: | sfigmomanometr |
Model: | MIT Elite |
Typ wyświetlacza: | LCD |
Ilość na paczkę: | 1 szt. |
Przeznaczenie: | Przedramię |
Rodzaj baterii: | AA |
Model: | Automatyczny |
Liczba baterii: | 4 |
Baterie w zestawie: | Tak |
Walizka: | Tak |
Gniazda pamięci: | 90 |
Pomiar tętna: | Tak |
Średnie ciśnienie tętnicze: | Tak |
Rozkurczowe ciśnienie kwi: | Tak |
Skurczowe ciśnienie krwi: | Tak |
Rozmiar mankietu: | 22 - 32 cm |
Funkcja pamięci: | Tak |
Data na zdjęciu: | Tak |
Potrzebujesz pomocy?
Jeśli potrzebujesz pomocy z Omron MIT Elite, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą
Instrukcje sfigmomanometr Omron

28 Maja 2024

26 Maja 2024

13 Maja 2024

6 Maja 2024
Instrukcje sfigmomanometr
- sfigmomanometr Beurer
- sfigmomanometr Sanitas
- sfigmomanometr Braun
- sfigmomanometr Panasonic
- sfigmomanometr Nedis
- sfigmomanometr Rossmax
- sfigmomanometr Citizen
- sfigmomanometr Homedics
- sfigmomanometr Scala
- sfigmomanometr Medisana
- sfigmomanometr Dr. Senst
- sfigmomanometr Perel
- sfigmomanometr Aponorm
- sfigmomanometr Microlife
- sfigmomanometr Innoliving
- sfigmomanometr Hama
- sfigmomanometr Boso
- sfigmomanometr Terraillon
- sfigmomanometr Sencor
- sfigmomanometr Etekcity
- sfigmomanometr Ecomed
- sfigmomanometr Veroval
- sfigmomanometr OBH Nordica
- sfigmomanometr Visomat
- sfigmomanometr Orava
Najnowsze instrukcje dla sfigmomanometr

24 Września 2024

21 Września 2024

31 Sierpnia 2024

31 Sierpnia 2024

24 Sierpnia 2024

19 Sierpnia 2024

30 Lipca 2024

30 Lipca 2024

8 Lipca 2024

8 Lipca 2024