Instrukcja obsługi Greisinger T-Logg 120 K


Przeczytaj poniżej 📖 instrukcję obsługi w języku polskim dla Greisinger T-Logg 120 K (8 stron) w kategorii Niesklasyfikowane. Ta instrukcja była pomocna dla 3 osób i została oceniona przez 2 użytkowników na średnio 4.5 gwiazdek

Strona 1/8
E39.0.1X.6C-01
GREISINGER electronic GmbH
D - 93128 Regenstauf, Hans-Sachs-Straße 26
℡  +49 (0) 9402 / 9383-0 +49 (0) 9402 / 9383-33 | info@greisinger.de
Data logger for standard signals
as of version 1.0 Operating Instruction T-Logg 120 ...
T-Logg 120 W - ...
T-Logg 120 K - ...
E39.0.1X.6C-01 Operating Instruction T-Logg 120 page 2 of 8
1 General
The logger T-Logg 120 ... is designed as a cost-efficient solution for monitoring standard signals. It enables an in-
dividual programming of the recording time. The last 16,000 measuring values can be stored in the memory.
In addition, the LCD-display indicates both the temperature measured at the moment and the operating status of
the logger.
2 Required accessory
The USB interface of your PC is used to program, start and read out the T-Logg.
For this following accessory is required:
Interface converter USB 100 or USB 100 SL for direct connection to the USB port of the PC,
cable length approx. 1m.
MINISOFT software (version 7.18 or later / free of charge) to start the logger and read out the logger data.
Notice: It’s also possible to use the T-Logg with the comprehensive software GSOFT 40K
(version 7.18 or later).
3 Safety instructions
This device has been designed and tested in accordance with the safety regulations for electronic devices.
However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and
special safety advises given in this manual will be adhered to when using the device.
1. Trouble-free operation and reliability of the device can only be guaranteed if the device is not subjected to any
other climatic conditions than those stated under ‘Specification’
To protect the battery the max. permissible storage and transport temperature of the device is +85°C.
2. General instructions and safety regulations for electric, light and heavy current plants, including domestic safety
regulations (e.g. VDE), have to be observed.
3. If device is to be connected to other devices (e.g. via PC) the circuitry has to be designed most carefully. Internal
connection in third party devices (e.g. connection GND and earth) may result in not-permissible voltages im-
pairing or destroying the device or another device connected.
4. If there is a risk whatsoever involved in running it, the device has to be switched off immediately and to be
marked accordingly to avoid re-starting.
Operator safety may be a risk if:
- there is visible damage to the device.
- the device is not working as specified.
- the device has been stored under unsuitable conditions for a longer time.
In case of doubt, please return device to manufacturer for repair or maintenance.
5. Warning:
Do not use this product as safety or emergency stop devices, or in any other application where failure of the
product could result in personal injury or material damage.
Failure to comply with these instructions could result in death or serious injury and material damage!
4 Connection
4.1 Interface connection
Assignment of the input jack
(front view to the pins):
interface [–]
not connected
interface [+]
E39.0.1X.6C-01 Operating Instruction T-Logg 120 page 3 of 8
4.2 T-Logg 120 W - ... : Assignment of the angle-type plug
Table of connections angle-type socket and jack
Inputs
Pin no. wire colour 0 ... X Volt 0 ... 20 mA 4 ... 20 mA
1 blue feeded through, signal + feeded through feeded through
2 red feeded through , signal - feeded through T-Logg
3 black feeded through T-Logg feeded through
4 yellow feeded through feeded through feeded through
Input signal: 0 ... X Volt Input signal: 0 ... 20 mA Input signal: 4 ... 20 mA
In the angle-type plug the male
contacts 1, 2 and 4 are directly
connected 1:1 with the socket.
The T-Logg (Rs = ~70 Ohm) is
located between the male con-
tact 3 (-) and the jack 3 (+).
In the angle-type plug the male
contacts 1, 3 and 4 are directly
connected 1:1 with the socket.
The T-Logg (Rs = ~70 Ohm) is
located between the male con-
tact 2 (-) and the jack 2 (+).
pin contact
Rs
+
_
1 2 3 4
1 2 3 4
jack
pin contact
Rs
+
_
1 2 3 4
1 2 3 4
jack
In the angle-type plug the 4 male
contacts are directly connected
1:1 with the socket.
The logger connection is on
contact 1 (signal+) and contact 2
(signal-).
If your transmitter assignments
for the 'signal+' and 'signal-' are
different, please do not forget to
adjust male contacts of your an-
gle-type plug accordingly:
To do so open the angle-type
plug and remove the red and the
blue wires entering the housing
from the coupling of the angle-
type plug.
Connect wires with the respec-
tive contacts, representing signal
+ (blue wire) and signal - (red
wire) at your transmitter.
If the 'signal'-line in your trans-
mitter is not assigned to contact
3, please do not forget to adjust
the logger-angle-type plug and
the external angle-type plug ac-
cordingly:
To do so open the logger-angle-
type plug and exchange the
black wire of contact 3 against
the wire of the contact repre-
senting the 'signal' in your trans-
mitter.
Then exchange and/or rewire the
two contacts in the angle-type
plug of your connecting cable.
If the 'signal/GND'-line in your
transmitter is not assigned to
contact 2, please do not forget to
adjust the logger-angle-type plug
and the external angle-type plug
accordingly:
To do so open the logger-angle-
type plug and exchange the red
wire of contact 2 against the wire
of the contact representing the
'signal/GND' in your transmitter.
Then exchange and/or rewire the
two contacts in the angle-type
plug of your connecting cable.
How to change the assignment of the angle-type plug:
- Dismantle the plug by pulling the adapter inset out of the case, using a screw-driver at the lateral groove.
- Change the assignment according to the notes at the respective input signal.
- Latch coupling insert in cover. You have a choice between 4 different orientations -
each of them spaced 90°.
- Please make sure to adjust the connecting cable, if the assignment of the angle-type plug has been
changed for the types 4-20mA and 0-20mA.
- Put on angle-type plug and connect plugs using the long screw delivered. (Do not forget seals).


Specyfikacje produktu

Marka: Greisinger
Kategoria: Niesklasyfikowane
Model: T-Logg 120 K

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Greisinger T-Logg 120 K, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Niesklasyfikowane Greisinger

Instrukcje Niesklasyfikowane

Najnowsze instrukcje dla Niesklasyfikowane

Stihl

Stihl HT 70 Instrukcja

15 Października 2024
Ernesto

Ernesto H14270 Instrukcja

15 Października 2024
Uniden

Uniden BT MIC KIT Instrukcja

15 Października 2024
Vaude

Vaude Omnis Bike 26 Instrukcja

15 Października 2024
ART

ART ProChannel II Instrukcja

15 Października 2024
Aukey

Aukey DR02J Instrukcja

15 Października 2024
Master Lock

Master Lock 653EURD Instrukcja

15 Października 2024
Roland

Roland FR-18 Diatonic Instrukcja

15 Października 2024