Instrukcja obsługi Global Water FL16


Przeczytaj poniżej 📖 instrukcję obsługi w języku polskim dla Global Water FL16 (37 stron) w kategorii Niesklasyfikowane. Ta instrukcja była pomocna dla 10 osób i została oceniona przez 2 użytkowników na średnio 4.5 gwiazdek

Strona 1/37
Global Water
800-876-1172
globalw.com
- 1 -
FL16 Flow Logger
Covers USB (FL16U) and Serial (FL16S) Versions
Global Water
Instrumentation, Inc.
11390 Amalgam Way
Gold River, CA 95670
T: 800-876-1172
Int’l: (916) 638-3429, F: (916) 638-3270
globalw@globalw.com
11/30/09
01-497
Global Water
800-876-1172
globalw.com
- 2 -
Congratulations on your purchase of a Global FL16 Flow Logger. This
instrument has been quality tested and approved for providing accurate
and reliable measurements. We are confident that you will find the logger
to be a valuable asset to your applications. Should you require
assistance, our technical staff will be happy to help.
Table of Contents
I. Checklist ▫▫▫▫▫ 3
II. Inspection ▫▫▫▫ 3
III. Warranty ▫▫▫▫▫ 3
IV. FL16 Specifications ▫ ▫ ▫ 4
V. Introduction ▫▫▫▫▫ 6
VI. Software Compatibility ▫▫▫ 7
VII. System Requirements ▫▫▫ 7
VIII. Software Installation ▫ ▫ ▫ ▫ 8
IX. Direct Connection ▫ ▫ ▫ 11
X. Modem Connection ▫▫▫▫ 12
XI. Main Screen ▫▫▫▫▫ 15
XII. On-Line Help Files ▫▫▫▫ 19
XIII. Get History ▫▫▫ ▫ ▫ 20
XIV. Setup Menus ▫▫▫▫▫ 21
XV. General Setup ▫▫▫▫▫ 22
XVI. Analog Setup ▫▫▫▫▫ 25
XVII. Flow Calculation Setup ▫▫▫ 27
XVIII. Programming ▫▫▫▫▫ 32
XIX. Calculations ▫▫▫▫▫ 33
XX. Modem Cable Diagram ▫▫▫ 34
XXI. Maintenance ▫▫▫▫▫ 35
XXII. Troubleshooting ▫▫▫▫ 35
XXIII. Customer Support ▫▫▫▫ 36
XXIV. CE Certification ▫▫▫▫ 37
Copyright © Global Water Instrumentation, Inc. 2007
Global Water
800-876-1172
globalw.com
- 3 -
I. Checklist:
a. FL16S (Serial) or FL16U (USB) Data Logger
b. FL16 Sewer Flow Level Sensor
c. FL16 Flow Logger Manual
d. FL16S comes with a RS-232 Serial Cable
e. FL16U comes with a USB Cable, Type A to B
f. Global Logger Interface Software on CDROM
II. Inspection:
a. Your Data Logger was carefully inspected and certified by our Quality
Assurance Team before shipping. If any damage has occurred during
shipping; please notify Global Water Instrumentation, Inc. and file a
claim with the carrier involved.
III. Warranty:
a. Global Water Instrumentation, Inc. warrants that its products are
free from defects in material and workmanship under normal use
and service for a period of one year from date of shipment from
factory. Global Water’s obligations under this warranty are
limited to, at Global Water’s option: (I) replace or (II) repair; any
products determined to be defective. In no case shall Global
Water’s liability exceed the products original purchase price.
This warranty does not apply to any equipment that has been
repaired or altered, except by Global Water Instrumentation, Inc.,
or which has been subject to misuse, negligence or accident. It
is expressly agreed that this warranty will be in lieu of all
warranties of fitness and in lieu of the warranty of
merchantability.
b. The warranty begins on the date of your invoice.
Global Water
800-876-1172
globalw.com
- 4 -
IV. FL16 Specifications:
Logger:
Power Supply: Two 9VDC Lithium batteries standard
8 VDC Min. to 24.0 VDC Absolute Maximum
Standby Current: 65uA Typical
Logging Current: 5mA Typical + sensor current
Operating Temperature Industrial, -40ºC to +85ºC (Battery may not apply)
Analog Sensor Inputs: 4-20mA
Resolution: 12-Bit, 4096 Steps
Level, temperature and battery voltage monitor
Sensor Warm-up Time: Programmable, 0-15 Seconds
Sample Modes: Fixed Interval: Programmable from 1 Sec. to >1 Year
High Speed: 10 Samples per second
Logarithmic Sample Rate (Approximation)
Exception (Log only on deviation from previous reading)
Storage Capacity: 81,759 Recordings of level, temp, battery voltage plus
date/time stamp
Communication Ports: FL16S: RS-232 4-pin circular connector
FL16U: USB Type B
Selectable Baud Rates:
9600, 19200, 28800, 38400, 57600, 115200
Housing: 1 7/8 diameter x 11 1/2 long, Stainless Steel and PVC
Weight: 450g (16oz), Logger only with batteries
Global Water
800-876-1172
globalw.com
- 5 -
Level Sensor:
Sensor Element: Silicone Diaphragm, Wet/Wet Transducer
Range: Level: 0-3 feet
Temperature: 32°-122°F (0°-50°C)
Linearity and Hysteresis: ±0.1% FS
Accuracy: Level: ±0.1% FS at constant temperature
±0.2% over 32° to 70°F range
Temperature: Smaller of 0.5°F or ±1% of reading
Overpressure: 2 x full scale
Burst Pressure: 10 x full scale
Resolution: Infinite (Analog)
Outputs: 4-20mA ±1mA at full scale
Optional Temperature: 0-10mA ±1mA FS
Supply Voltage: 10-36VDC
Current Draw: Sum of sensor outputs.
Warm Up Time: 10mS Min, 3 sec. recommended
Operating Temperature: (Not Frozen) to +185°F
Compensated Range: 32° to 70°F submerged, automatic barometric
compensation
Sensor Housing:
Material: 316 SS
Size: 9” long x 1.0” diameter
Weight: 250g (9oz)
Cable:
Conductors: 4 each 22 AWG
Jacket Material: 87A shore hardness Polyurethane
Optional jacket: Fluorinated Ethylene Propylene (FEP) Teflon
Cable O.D.: 7.8mm (0.307”)
Vent tube: HD Polyethylene
Shield: Aluminum Mylar
Temperature range: -30 to 85°C (-22 to 185°F)
Weight: ~65g/m (0.7 oz/ft)
Global Water
800-876-1172
globalw.com
- 10 -
6) Two drivers are sometimes required so the Hardware Wizard repeats the exact
same process again. As before; a: do not let the Windows Update do a search, b:
tell Windows to search from a specific location instead of automatically installing
software, c: tell Windows to search from removable media (CDROM).
7) The installation is complete.
Click Finish.
8) To confirm the installation, look in the
Device Manager and find the USB Serial
Port. The COM setting following that will
show the virtual COM port that has been
automatically assigned to the driver. As
you will see in the following sections, this
COM port will be automatically detected
by the Global Logger software so it is not
necessary to note which port is being
used by the driver.


Specyfikacje produktu

Marka: Global Water
Kategoria: Niesklasyfikowane
Model: FL16

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Global Water FL16, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Niesklasyfikowane Global Water

Instrukcje Niesklasyfikowane

Najnowsze instrukcje dla Niesklasyfikowane

SmallRig

SmallRig VB99 SE Instrukcja

29 Stycznia 2025
QSC

QSC TSC-8 Instrukcja

29 Stycznia 2025
XTRARM

XTRARM XTR-FS5130 Instrukcja

29 Stycznia 2025
Sunding

Sunding SD-581G Instrukcja

29 Stycznia 2025
XTRARM

XTRARM XTR-FS5070 Instrukcja

29 Stycznia 2025
XTRARM

XTRARM XTR-FS2010 Instrukcja

29 Stycznia 2025
OSEE

OSEE TX402 Instrukcja

29 Stycznia 2025