Instrukcja obsługi Extech HT30


Przeczytaj poniżej 📖 instrukcję obsługi w języku polskim dla Extech HT30 (6 stron) w kategorii termometr. Ta instrukcja była pomocna dla 3 osób i została oceniona przez 2 użytkowników na średnio 4.5 gwiazdek

Strona 1/6
User's Manual
Heat Stress WBGT Meter
Model HT30
2 HT30-en-GB_V3.6 4/16
Introduction
Congratulations on your purchase of Heat Stress WBGT (Wet Bulb Globe Temperature) Meter. This
meter measures and displays Heat Stress Index (WBGT), which is how hot it feels when humidity is
combined with temperature, air movement, and direct or radiant sunlight. Black Globe Temperature
(TG) monitors the effects of direct solar radiation on an exposed surface. It also measures air
temperature (TA) and Relative Humidity (RH). Additional features include: selectable units of °F/°C,
Auto power off with override, and a built-in RS-232 interface with optional 407752 Windows®
software. This meter is shipped fully tested and calibrated and, with proper use, will provide years of
reliable service.
Meter Description
1. Black Globe temperature sensor
2. RH and Temperature sensors with protective cover
3. LCD display
4. /SET
5. NEXT
6. MODE/
7. RS-232 interface
8. Battery compartment (on rear)
DISPLAY
Symbol Function
WBGT Wet Bulb Globe Temperature
TG Black Globe Temperature
TA Air Temperature
RH% Relative Humidity
OUT Indoor (no sun)
IN Outdoor (in full sun)
C/F Celsius/Fahrenheit
Low Battery indicator
3 HT30-en-GB_V3.6 4/16
Meter Operation
1. Press the /SET button to power the instrument on/off.
2. Slide down the protective sensor cover before taking measurements.
3. Press the MODE/button to select the desired display mode: Wet Bulb Globe Temperature
Index (WBGT), Air Temperature (TA), Black Globe Temperature (TG), or Relative Humidity
(RH). An icon will appear in the display indicating the current selection.
4. To select the preferred temperature unit (°F or °C), simultaneously press and release the
MODE/and NEXT buttons. The F or C icon will appear in the display.
5. The meter measures WBGT index with (IN) or without (OUT) direct sun exposure. Hold down
the MODE/ for more than 1 second to alternate between settings. The IN or OUT icon will
appear in the display.
Alarm Setting
The HT30 has a user settable heat stress alarm. The HT30 will emit an audible beep when the heat
stress WBGT reaches the level previously set by the user. The alarm will continue beep until the
measured WBGT temperature falls below the alarm level set by the user, or the meter is turned off.
The setting range is from 68.0 to 99.0°F (20.0 to 37.2°C).
1. To enter the alarm setup mode, hold down the /SET button for more than 2 seconds when
turning the meter on.
2. The current alarm setting will be displayed with the most significant digit blinking.
3. Use the MODE/button to increment the digit.
4. To set the next digit, momentarily press the NEXT button.
5. After all of the digits are set, press and hold the NEXT button for 2 seconds to return to the
measuring mode.
NOTE: The OUT icon will appear in the display If the new setting is out of the setting range of the
alarm.


Specyfikacje produktu

Marka: Extech
Kategoria: termometr
Model: HT30

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Extech HT30, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje termometr Extech

Extech

Extech HT30 Instrukcja

29 Września 2024
Extech

Extech EN300 Instrukcja

29 Września 2024
Extech

Extech IR267 Instrukcja

11 Września 2024

Instrukcje termometr

Najnowsze instrukcje dla termometr

Auriol

Auriol H14269 Instrukcja

15 Października 2024
Medel

Medel ECOtemp Instrukcja

9 Października 2024
Medel

Medel No Contact Plus Instrukcja

9 Października 2024
Medel

Medel Flexo Instrukcja

9 Października 2024
Owon

Owon TI300 Instrukcja

8 Października 2024
Innoliving

Innoliving INMD-DMT-437 Instrukcja

8 Października 2024