Instrukcja obsługi Beurer GL 50 evo


Przeczytaj poniżej 📖 instrukcję obsługi w języku polskim dla Beurer GL 50 evo (44 stron) w kategorii miernik glukozy we krwi. Ta instrukcja była pomocna dla 3 osób i została oceniona przez 2 użytkowników na średnio 4.5 gwiazdek

Strona 1/44
3-IN-1 BLOOD GLUCOSE MONITOR
Step by step
GL 50 evo
EN
Beurer GL50 evo 3
1 Getting to know your device
.........................................................................................
4
1.1 Included in the delivery and accessories
...............................................................................
5
1.2 Replacements
........................................................................................................................
6
1.3 Functions of the device
..........................................................................................................
6
1.4 Signs and symbols
................................................................................................................
7
2 Warnings and safety notes
............................................................................................
8
3 Description of device and accessories
.................................................................
11
3.1 Blood glucose monitor .........................................................................................................11
3.3 USB cover
............................................................................................................................
12
3.4 Display symbols
...................................................................................................................
12
3.5 Test strips
............................................................................................................................
13
4 Initial use and basic settings
......................................................................................
14
4.1 Charging the battery
............................................................................................................
14
4.2 Making and changing basic settings
...................................................................................
15
5 Taking measurements
....................................................................................................
20
5.1 Preparing to take a blood sample
........................................................................................
20
5.2 Preparing the lancing device for taking a sample
................................................................
21
5.3 Taking a blood sample and measuring the glucose level
....................................................
22
5.4 Reading the result and marking measurements
..................................................................
24
5.5 Post-processing and disposal
.............................................................................................
25
5.6 Evaluating measured blood glucose values
........................................................................
26
5.7 Functional check using control solution
..............................................................................
28 6 M easureMent MeMory ....................................................................................................30
6.1 Displaying individual values
.................................................................................................
31
6.2 Displaying average blood glucose values for marked values
..............................................
31
6.3 Evaluating measured values on a PC
..................................................................................
32
7 Storing, maintaining and disinfecting the device
.............................................
32
7.1 Maintenance
........................................................................................................................
33
7.2 Disinfection
..........................................................................................................................
33
8 What if there are problems?
........................................................................................
34
9 Technical specifications
................................................................................................
36
10 Comparison of measured values with laboratory values
.............................
37
11 Usage limits for specialist personnel from the healthcare sector
..........
39
12 Warranty/service ............................................................................................................... 41
Contents
4 Beurer GL50 evo
1 G ettinG to know your device
Dear customer,
Thank you for choosing a product from our range. Our name stands for high-quality, thoroughly
tested products for applications in the areas of heat, weight, blood pressure, blood glucose, body
temperature, pulse, gentle therapy, massage and air.
Please read these instructions for use carefully and keep them for later use. Be sure to make them
accessible to other users and observe the information they contain.
With kind regards,
Your Beurer team.
Getting to know your device
The GL50 evo blood glucose monitor is intended for fast and simple blood glucose measurement of
fresh capillary blood samples, either for self-testing or in a clinical environment by trained personnel.
It enables you to measure your blood glucose quickly and easily, store the measured values and
display the average of all measured values, thereby providing optimum assistance for monitoring your
diabetes. The test is performed exclusively externally (IVD).
The backlit display shows measured values clearly. The user-friendly design with handy test strips
and the easy operation via a single scroll wheel guarantee simple, yet reliable measurements.
The device can be connected directly to a PC using the integrated USB connection. You can evaluate
the measured values on your PC using special software (in English and German language) and use
the results to monitor your blood glucose values.
Alternatively, you can transfer the measured values to your smartphone/tablet using the optional
Bluetooth® cap.
Beurer GL50 evo 5
1.1 Included in the delivery and accessories
Check that the set packaging has not been tampered with and make sure that all contents are present.
Before use, ensure that there is no visible damage to the device or accessories and that all packaging
material has been removed. If you have any doubts, do not use the device and contact your retailer
or the specified Customer Service address.
A B
EDC
A “3-in-1” monitor: lancing device, plug-in USB and monitor
B USB cover
C 10 test strips
D 10 sterile lancet needles (5 x 28G blue, 5 x 33G purple)
E 1 practical case
These instructions for use, additional information
The blood glucose monitor (A), test strips (C) and additionally available control solutions have been
specially designed to complement each other. For this reason, use only test strips (C) and control
solutions that have been approved for this blood glucose monitor (A).
Note
Use original manufacturer accessories only.
6 Beurer GL50 evo
1.2 Replacements
You can also obtain test strips, control solution and lancets without a prescription.
Item REF
50 test strips REF 464.15
50 test strips, individually film packed REF 464.17
100 test strips REF 464.13
Control solution LEVEL 3 and 4 REF 464.16
100 soft touch lancets 33G REF 457.24
100 lancet needles 28G REF 457.01
100 safety lancets REF 457.41
200 safety lancets REF 457.42
1.3 Functions of the device
This device is intended for measuring the blood glucose content in human blood. It is also suitable
for self-testing at home.
The monitor enables you to quickly and simply:
Measure your blood glucose level
Display, mark and save measured values
Output the measured values acoustically
Display the average of the measured blood glucose values from the last 7, 14, 30 and 90 days
Display the average of the marked measured blood glucose values from the last 7, 14, 30 and 90 days
Set the timer and alarm
Set the time and date
Evaluate the saved measured values on a PC using special software
Transfer the saved measured values to a smartphone/tablet using the optional Bluetooth
®
cap
Transfer the saved measured values to a smartphone/tablet using the optional NFC cap
The monitor also includes the following monitoring functions:
Warning that test strip is insuciently filled
Warning in the event of unsuitable temperatures
Ketone warning
State of charge display when rechargeable battery is low.
Warning
Do not use the device to diagnose diabetes; it is intended for regular monitoring only
Consult your GP with regard to insulin doses.
Beurer GL50 evo 7
1.4 Signs and symbols
The symbols on the packaging and type plate of the monitor and accessories represent the following:
In vitro diagnostic device Manufacturer
Serial number Observe the instructions for use
Temperature limit
+2°C to +30°C
Not for re-use/for single use
only
Use by
Green dot (Der Grüne Punkt):
German dual waste collection
system
18 M
Maximum shelf life after initial
opening in months Contents sucient for <n> tests
L O T
Batch designation /
Art.-Nr.
Order number
Sterilised by radiation (lancets) Unit of measurement for blood
glucose value
Warning, see accompanying
documents Biohazard, risk of infection
Medical device (MDR Symbol)
CE labelling
This product satisfies the
requirements of the applicable
European and national
directives.
Rechargeable battery
In the instructions for use, the symbols represent the following:
Warning
Warning instruction indicating a risk of injury or damage to your health/your patient's health.
Important
Safety note indicating possible damage to the device/accessory.
Note
Note on important information.
Beurer GL50 evo 9
A haematocrit value (proportion of red blood cells) between 20% and 60% has no significant influence
on the measurement results.
A very high or low haematocrit value (proportion of red blood cells) may lead to incorrect
measurements. In the event of a very high haematocrit value (above 60%), the displayed blood
glucose value may be too low; in the event of a very low haematocrit value (below 20%), it may be
too high. Consult your GP if you do not know your haematocrit value.
Do not use the test strips to measure blood glucose values on newborns.
Do not use NaF or potassium oxalate anticoagulants to prepare for venous blood samples.
Do not test any severely ill patients using this device.
Use fresh whole blood only. Do not use serum or plasma.
Use capillary blood without squeezing the penetration area. Squeezing the area causes the blood to
be diluted with tissue fluid and this may lead to an incorrect measurement.
Do not use the test strips above an altitude of 7010 metres.
Very high levels of humidity may influence the test results. Relative humidity of more than 90% may
lead to inaccurate results.
Note
The Beurer GL50 evo measuring system is intended for measuring capillary and venous whole blood.
Storage and maintenance
Warning
Store the blood glucose monitor and its accessories out of reach of small children and pets. Small
parts, such as lancet needles, cover caps or test strips may be life-threatening when swallowed. If
swallowed, seek medical attention immediately.
The test strip box contains desiccant, which may irritate the skin or eyes when inhaled or swallowed.
Keep the box out of the reach of children.
The monitor is made from precision and electronic components. The accuracy of the measured values
and service life of the device depend on its careful handling:
Protect the device and its accessories from impacts, humidity, dirt, marked temperature fluctuations
and direct sunlight. Do not store the device, test strips and control solution in your vehicle, in the
bathroom or in a cooling appliance.
Do not drop the device.
Repairs
Note
Do not open the device. Failure to comply with this instruction will void the warranty.
Do not repair the device yourself. Proper operation can no longer be guaranteed in this case.
Please contact Customer Services for repairs.
10 Beurer GL50 evo
Disposal
Warning
It is essential to comply with the generally applicable safety precautions for handling blood when
disposing of materials from the blood glucose monitor. Dispose of all blood samples and materials
with which you or your patients come into contact correctly in order to prevent injury and infection
of other persons.
After use, dispose of test strips and lancets in a puncture-proof container.
Note
Rechargeable batteries must not be disposed of with household waste. As a consumer, you are
required by law to recycle used rechargeable batteries. You can recycle your old rechargeable batteries
at public collection points in your community or wherever rechargeable batteries of the relevant type
are sold.
For environmental reasons, do not dispose of the device in the household waste at the end
of its service life. Please dispose of the device in accordance with EC Directive WEEE
(Waste Electrical and Electronic Equipment). If you have any questions, please contact the
local authorities responsible for waste disposal.
Notes on handling batteries
If your skin or eyes come into contact with fluid from the battery cell, flush out the aected areas
with water and seek medical assistance.
Choking hazard! Small children may swallow and choke on batteries. Store batteries out of the reach
of small children.
If a battery has leaked, put on protective gloves and clean the battery compartment with a dry cloth.
Risk of explosion! Never throw batteries into a fire.
Do not disassemble, split or crush the batteries.
Only use chargers specified in the instructions for use.
Batteries must be charged correctly prior to use. The instructions from the manufacturer and the
specifications in these instructions for use regarding correct charging must be observed at all times.
Rechargeable battery disposal
The empty, completely flat rechargeable batteries must be disposed of using specially designated
collection boxes, recycling points or electronics retailers. You are legally required to dispose of the
batteries.
Beurer GL50 evo 11
3 d escription of device and accessories
3.1 Blood glucose monitor
An overview of the monitor
1 2 3
1 Lancing device
2 Plug-in USB
3 Monitor
Front Side
1 Display
2 Slot for test strips
1
2
1 Scroll wheel
2 ON/OFF switch
3 Tensioning slider
1
2
3
3.2 Lancing device and lancet needles
1 Cap
2 Protective lancet disc
3 Sterile lancet needle
4 Lancet holder
5 Dial for setting dierent
penetration depths
6 Tensioning slider
7 Trigger
23457
6
1
12 Beurer GL50 evo
3.3 USB cover
If you would like to use the blood glucose monitor without
the integrated lancing device, you can use the supplied
USB cover in place of the lancing device.
3.4 Display symbols
1
4
2
3
6
7
5
memory
1 State of charge
2 Memory space number
3 Blood sugar unit mg/dL or mmol/L
4 Time
5 Symbol for marking measurements
6 Date
7 Measured value display
Note
The monitor is supplied with the following basic settings:
Beep on
Bluetooth® o
NFC o
Ketone warning on
Language: German
Warning
Ensure that you are using the device with the correct blood glucose unit (either mg/dL or mmol/L)
setting for you. If in doubt, consult your GP.
Beurer GL50 evo 13
3.5 Test strips
Front Rear
1
2
3
1 Gap for blood input
2 Grip area
3 Contacts
You can identify the rear by the contact tracks.
Insert the test strip into the device
so that the contacts are pointing
inside the slot.
Make sure that the front of the test
strip is facing you.
Note
Read carefully the following information on handling and storing your test strips. The test strips will
only provide accurate measurements if all information is followed.
Warning
Use each test strip only and for patient only!once one
Handling test strips
Note
Securely close the test strip box immediately after taking out a test strip.
Do not use the test strips if they have expired. The use of expired test strips may lead to incorrect
measurements. The expiry date is located next to the hourglass symbol on the box or on the
respective film packaging of the individual test strips.
The test strips expire 18 months after the box is opened (note down the expiry date (date of opening
plus 18 months
18 M
) on the label. In the event of overlap, the shelf life is limited to the expiry date (see
14 Beurer GL50 evo
date next to the hourglass symbol ). This does not apply for individual test strips, which are to be
used immediately after opening the film packaging.
Discontinue use of the test strips if one of the two expiry dates ( /
18 M
) has passed.
You can touch any part of the test strip with clean, dry hands.
Use the test strip for measurement immediately after removing it from the box/film packaging.
Do not bend, cut or otherwise modify the test strips.
Do not use test strips that have come into contact with fluids.
Storing test strips
Note
Keep the test strips in a cool, dry place above +2°C and below +30°C. Do not expose the test strips
to direct sunlight or heat. Do not store in your vehicle, in the bathroom or in a cooling appliance.
Permitted relative air humidity below 90%.
The test strips must be stored in the original box/unopened film packaging never use other
containers.
4 i nitial use and basic settinGs
In order to activate the device, move the ON/OFF switch into the “ON” position.
on
o
The ON/OFF switch is used as a keylock for the scroll wheel.
Information regarding initial use
Before putting the blood glucose monitor into operation for the first time, you must fully charge the
rechargeable battery. To charge the rechargeable battery, proceed as directed in the following chapter.
Before the initial use, no battery symbol will be displayed during charging.
4.1 Charging the battery
1
Remove the lancing device from the blood glucose monitor.
2
Connect the blood glucose monitor to a computer via the USB port.
Leave the blood glucose monitor to charge for at least 2 hours.
3
Once you have fully charged the rechargeable battery of the blood glucose monitor, disconnect
the blood glucose monitor from the computer.
4
Place the lancing device back on the blood glucose monitor.
After charging the battery, wait 30 minutes before carrying out a blood glucose measurement.
Beurer GL50 evo 15
4.2 Making and changing basic settings
1
Setting the date and time
Note
You must set the date and time. Otherwise, you will not be able to save your measured values
correctly with a date and time and access them again later.
The time can be displayed in 12-hour format or 24-hour format.
1. Switch on the blood glucose monitor by briefly pressing and holding the scroll wheel.
2. Rotate the scroll wheel until the arrow is pointing towards
“Settings”, then press the scroll wheel.
• All values
• Average
• Settings
Back
3. Rotate the scroll wheel until the arrow is pointing towards “Date/
time”, then press the scroll wheel.
• Timer
• Alarm
• Date/time
• Bluetooth on/o
4. Set the year by rotating the scroll wheel, then press the scroll
wheel.
Year 2018
Month 10
Day 21
24h/12h mode 24h
5. Use the same procedure to set the month, day, 24 h or 12 h mode, hour and minute.
6. To save the settings, rotate the scroll wheel until the arrow is
pointing towards “Save” and confirm by pressing the scroll
wheel.
24h/12h mode 24h
Hour 12
Minute 00
Save
Beurer GL50 evo 17
3
6. In order to start the alarm, rotate the scroll wheel to “On” and
press the scroll wheel.
7. In order to turn o the alarm, rotate the scroll wheel to “O”
and press the scroll wheel.
alarm
on
4
Switching Bluetooth® or NFC on/o
You can switch on Bluetooth®/NFC on the blood glucose monitor.
1. Switch on the blood glucose monitor by briefly pressing and
holding the scroll wheel.
2. Rotate the scroll wheel until the arrow is pointing towards
“Settings, then press the scroll wheel.
• All values
• Average
• Settings
Back
3. Rotate the scroll wheel until the arrow is pointing towards
“Bluetooth on/o” or “NFC on/o” and press the scroll wheel.
• Alarm
• Date/time
• Bluetooth on/o
• NFC on/o
4. Rotate the scroll wheel until the arrow is pointing towards
“Bluetooth on”, then press the scroll wheel. Bluetooth
® is now
switched on. Proceed in the same manner to switch on NFC.
• Bluetooth on
• Bluetooth o
Back
18 Beurer GL50 evo
5
Switching the beep on/o
On the blood glucose monitor, you can switch on the beep, switch o the beep or set a result
beep.
1. Switch on the blood glucose monitor by briefly pressing and holding the scroll wheel.
2. Rotate the scroll wheel until the arrow is pointing towards
“Settings, then press the scroll wheel.
• All values
• Average
• Settings
Back
3. Rotate the scroll wheel until the arrow is pointing towards
“Beep on/o”, then press the scroll wheel.
• Bluetooth on/o
• NFC on/o
• Beep on/o
• Ketone on/o
4. Use the scroll wheel to select your desired setting (beep on,
beep o or result beep) and confirm by pressing the scroll
wheel.
• Beep on
• Beep o
• Result beep
Back
Note
The result beep transmits the measurement as an acoustic beep signal.
Example: measured value 103
Start signal
Beep beep
Number 1
Short beep
Number 2
Long beep (for 0)
Number 3
3 x short beep
The beep signal is repeated 2 times and can be interrupted by rotating the scroll wheel.
6
Switching the ketone warning on/o
On the blood glucose monitor, you can switch a ketone warning on and o.
1. Switch on the blood glucose monitor by briefly pressing and
holding the scroll wheel.
2. Rotate the scroll wheel until the arrow is pointing towards
“Settings, then press the scroll wheel.
• All values
• Average
• Settings
Back
3. Rotate the scroll wheel until the arrow is pointing towards
“Ketone on/o”, then press the scroll wheel.
• NFC on/o
• Beep on/o
• Ketone on/o
• Turn display
Beurer GL50 evo 19
6
4. Use the scroll wheel to select your desired setting (warning on
or warning o) and confirm by pressing the scroll wheel.
• Ketone
• Warning on
• Warning o
Back
7
Turning the display
You can turn the display on the blood glucose monitor.
1. Switch on the blood glucose monitor by briefly pressing and
holding the scroll wheel.
2. Rotate the scroll wheel until the arrow is pointing towards
“Settings, then press the scroll wheel.
• All values
• Average
• Settings
Back
3. Rotate the scroll wheel until the arrow is pointing towards “Turn
display”, then press the scroll wheel.
• Beep on/o
• Ketone on/o
• Turn display
• Language
8
Selecting the menu language
The following menu languages are available: English, German, French, Spanish, Italian and
Turkish.
1. Switch on the blood glucose monitor by briefly pressing and
holding the scroll wheel.
2. Rotate the scroll wheel until the arrow is pointing towards
“Settings, then press the scroll wheel.
• All values
• Average
• Settings
Back
3. Rotate the scroll wheel until the arrow is pointing towards
“Language”, then press the scroll wheel.
• Ketone on/o
• Turn display
• Language
Back
4. Use the scroll wheel to select your desired menu language and
confirm by pressing the scroll wheel.
• Language
Deutsch D
English G
FrançaisF
20 Beurer GL50 evo
5 t akinG MeasureMents
Warning
If the protective disc on a lancet needle has already been removed, do not use the lancet needle.
If you drop the lancing device with an inserted lancet needle, carefully pick it up and dispose of the
lancet.
Important
Use the lancing device only with lancet needles from the same manufacturer. Using other lancet
needles may prevent the lancing device from working properly.
If you are using a third-party lancing device, please read the accompanying instructions for use.
5.1 Preparing to take a blood sample
1
Choose a part of the body from which to take a blood sample
Using the lancing device, you can take a blood sample from the fingertip. To make the procedure
as painless as possible, do not take samples directly from the centre of the fingertip, but
slightly to either side.
Warning
In the event of suspected hypoglycaemia: take blood from the fingertip only. Reason:
changes to blood glucose levels can be detected quickly in blood samples taken from the
fingertip.
2
Prepare all parts
Prepare the following items: GL50 evo measuring device (A), test strip box or test strips in
film packaging (C) and sterile lancet needles (D).
Make sure that the ON/OFF switch is in the “ON” position.
3
Wash your hands
Wash your hands with soap and warm water before taking a blood sample. This not only
ensures optimal hygiene but also encourages good blood circulation at the puncture area on
the finger. Dry your hands carefully.
Warning
If you have used alcohol for cleaning, ensure that the area has fully dried prior to measuring.
Beurer GL50 evo 21
5.2 Preparing the lancing device for taking a sample
1
Remove the cap
Securely hold the monitor with one hand on the lancing device cover.
With your other hand, remove the cap from the monitor’s lancing device.
2
Insert the lancet needle
Insert a sterile lancet needle into the lancing device.
Note
Your starter set contains lancet needles in 2 dierent sizes. If you are
unable to take an adequate blood sample using the smaller needles
(purple, 33G), please use the slightly larger needles (blue, 28G).
Push firmly on the lancet until it audibly engages and it cannot be
pushed further into the holder.
3
Remove the protective lancet disc
Remove the protective lancet disc by bringing it into a horizontal position
and taking it o. Retain the protective disc for the safe disposal of the
used lancet needle after taking a blood sample.
4
Place the cap onto the lancing device. Make sure that the curved part
of the cap fits on the curved part of the lancing device. Press firmly on
the cap until it audibly engages.
5
Select the penetration depth
You can set seven dierent penetration depths on the lancing device
using the dial with printed numbers.
1 to 2: soft or thin skin
3 to 5: normal skin
6 to 7: thick or calloused skin
Turn the dial until the required bar is in the centre of the black marking.
6
Tension the lancing device
Pull back the slider in the direction of the arrow (on the image: to the
right) until it stops and then release. The slider automatically springs
back into position. The lancing device is now tensioned.
22 Beurer GL50 evo
5.3 Taking a blood sample and measuring the glucose level
Warning
Change the puncture area each time you take a measurement, e.g. use a dierent finger or the other
hand. Repeatedly using the same area may cause inflammation or scarring.
If the cap is not in place, there is a risk of injury from the exposed lancet.
Do not squeeze your finger to obtain a larger drop of blood. If squeezed, the blood is diluted with
tissue fluid and this may lead to an incorrect measurement.
Please note that insucient blood circulation at the puncture area, e.g. caused by cold temperatures
or illness, can lead to incorrect results.
Important
Do not apply any blood samples or control solutions to the test strip before inserting it in the monitor.
1
Prepare the test strip
Take a test strip from the box/film packaging and immediately close it again. Use the test strip
within three minutes of removal.
2
Insert test strip
Take the monitor in your left hand. Hold the monitor so
that the display is facing you.
Insert the test strip into the slot on the rear end of the
blood glucose monitor with the contacts first. Make sure
that the front of the test strip is facing you. You can touch
any part of the test strip with clean, dry hands.
3
The device switches on automatically
After you have inserted the test strip, the blood glucose
monitor switches on automatically and the Beurer logo,
the date and the time are briefly displayed.
The measuring device is ready to measure as soon as
the test strip animation is displayed.
Important
If segments of the display are missing, stop using the
device and immediately contact Customer Services.
Beurer GL50 evo 23
4
Lancing to take a blood sample
The lancing device can now be used to take a blood
sample. Make sure that the blood remains as a droplet
and is not spread.
Blood sample from the fingertip
Firmly position the lancing device slightly to the side of
the centre of the fingertip. Press the trigger. Remove the
lancing device from the finger. A round drop of blood of
at least 0.6 microlitres (corresponds to approx. 1.4 mm
in diameter, original size: ) must have formed.
Please also note the following:
If the blood glucose test results do not match how you feel, carry out another test using blood
from your fingertip.
DO NOT change your treatment purely on the basis of a measurement that was carried out
using blood taken from an alternative area. Carry out another test with blood from your fingertip
in order to confirm the test result.
If you often fail to notice that you have a low blood glucose level, carry out a test using blood
from your fingertip.
5
If necessary, repeat the lancing process
If you do not obtain sucient blood, repeat the lancing process with a greater penetration
depth in a dierent area.
6
Apply blood to the test strip
Turn the monitor by 180°. Hold the blood input gap
(at the tip of the test strip) to the drops of blood until the
gap is completely lled and the monitor in the display
starts counting backwards.
Do not press the penetration area to the test strip.
The blood must not be spread. The blood is sucked into
the gap.
OK
Note
Error message Error 2” appears on the display if the gap was not correctly and suciently filled
with blood. Repeat the measurement using a new test strip and a greater penetration depth.
24 Beurer GL50 evo
Note
Do not apply blood to the sides of the test strips
Do not add blood later if the device does not start measurement. Remove the test strip and end this
test. Use a new test strip.
If the test strip has already been inserted into the device but no blood is added to the test strip within
two minutes, the device switches itself o. Briefly remove the test strip and reinsert it so that the
device automatically switches itself back on.
Contact Customer Services if you are unable to fill the test strip with blood correctly.
5.4 Reading the result and marking measurements
Reading the result
As soon as the blood input gap is filled with sucient
blood, the device performs the blood glucose
measurement. The monitor counts down for approx. five
seconds. The measurement is then shown on the display.
MEMORY
MEMORY
Read your measured value. Check again that you have read the result correctly. The underline must
be below the measured value, otherwise you need to turn the monitor by 180°. For explanations of
and measures regarding the measured values, see “5.6 Evaluating measured blood glucose values”
on page 26. If an error message is displayed, read chapter “8. What if there are problems?on
page 34.
Marking the measurement
You have the following options for marking measured values:
ONo marking
Before meal
After meal
General marking (e.g. after exercising).
Marking measured values enables you, your GP or diabetes consultant to better monitor your blood
glucose values. For example, you can display the average values of all measurements taken before
a meal.
Beurer GL50 evo 25
To do this, proceed as follows:
1
Following the measurement, the display will briefly show the measured value and then the menu
in which you can mark the measured value. It is not possible to change the marking afterwards.
2
Select the desired marking using the scroll wheel.
O No marking”
“Before meal”
After meal”
“General marking”
Confirm the marking by pressing the scroll wheel.
3
The selected marking is now allocated to the measured value and saved to the memory.
5.5 Post-processing and disposal
1
Remove test strip
Remove the test strip from the device and carefully dispose of it in
accordance with the applicable regulations to avoid infecting others.
2
Remove the cap
Carefully remove the cap from the lancing device.
3
Stick protective disc on needle
Place the retained protective disc at on a hard surface. Stick the
tip of the needle into the protective disc (image 1) so the needle is
covered (image 2).
Image 1 Image 2
26 Beurer GL50 evo
4
Remove and dispose of lancet needle
Press the trigger again so that you can grip the shaft. Carefully remove
the lancet needle from the lancing device and dispose of the lancet in
a puncture-proof container.
Carefully dispose of all blood samples and materials that came into
contact with you or your patients. This prevents injuries and the
infection of others.
5
Replace cap
Place the cap back on the device.
5.6 Evaluating measured blood glucose values
Your blood glucose monitor can process measured values between 20 and 630 mg/dL (1.1 and
35.0 mmol/L). A warning message is displayed in the event of measured values lower than 20 mg/dL
(1.1 mmol/L) and measured values higher than 630 mg/dL (35.0 mmol/L).
Warning
If you suspect that the blood glucose results are incorrect, first repeat the test and, if necessary,
perform a functional test using control solution. Seek medical advice if dubious results persist.
Seek medical attention immediately if your symptoms do not correspond to your measured blood
glucose values and you have followed all instructions for the Beurer GL50 evo blood glucose monitor.
Do not ignore symptoms of too high/low blood glucose levels. Always seek medical attention!
Blood glucose values
The following tables list blood glucose values based on the STANDARDS OF MEDICAL CARE IN
DIABETES2016 from the ADA (American Diabetes Association).
Time of the blood glucose
measurement
Normal blood
glucose values
Increased risk of
diabetes (prediabetes)*
Diabetes
On an empty stomach
(fasting plasma glucose)
Below 100mg/dL
Below 5.6 mmol/L
100 125 mg/dL
5.6 – 6.9 mmol/L
≥ 126 mg/dL
≥ 7.0 mmol/L
Two hours after an oral
glucose tolerance test
(consumption of 75g)
Below 140 mg/dL
Below 7.8 mmol/L
140 199 mg/dL
7.8 – 11.0 mmol/L
≥ 200 mg/dL
≥ 11.1 mmol/L
*The risk increases continually, beginning with values below the lower limit of the range and increasing disproportionately towards
the upper limit of the range.
28 Beurer GL50 evo
Display ActionsBlood glucose
Very high blood
glucose level
Above 630 mg/dL
(over 35.0 mmol/L)
Take another measure-
ment using a new test
strip. If the display is
the same as before:
seek medical assistance
immediately.
5.7 Functional check using control solution
The control solution is used to test the entire blood glucose monitoring system. This helps to determine
whether the monitor and the test strips are working optimally together and whether the test is being
performed correctly.
Perform the control solution test if you suspect that the blood glucose monitor and/or the test strips
could be faulty or if you have repeatedly measured unusual blood glucose values. Also test the monitor
if it has been dropped or is damaged. The control solution is available separately. For the control
solution test, please observe the additional notes in the instructions for use for the control solution.
Important
Do not use third-party control solution. Correct functioning of your monitor can only be tested using
Beurer LEVEL3 + LEVEL4 control solutions.
Control solution measurements: when using the device, specialist personnel must follow statutory
guidelines.
Do not apply any blood samples or control solutions to the test strip before inserting it in the monitor.
Performing a functional test using control solution
Warning
To obtain correct results, the measuring device, the test strip and the control solution must be the
same temperature. For the “Functional test using the control solution”, the temperature is to be
between 20°C and 26°C.
1
Insert test strip
Hold the monitor so that the display is facing you. Insert a test strip into the slot on the monitor
with the contacts first. Make sure that the front of the test strip is facing you (see section “3.5
Test strips”).
Beurer GL50 evo 29
2
Activate control mode
Once you have inserted the test strip, the blood glucose monitor
switches on automatically and briefly shows the initial display. Once
the initial display has disappeared, quickly rotate the scroll wheel all
the way from the left to the right until “CONTROL appears at the top
of the animation display.
IMPORTANT: Control solutions and blood react to temperature influences in dierent ways.
It is therefore of vital importance that control solution measurement is always performed in
control solution mode. If this mode is not used, results may be obtained that are outside the
target range.
3
Drip control solution on surface
Choose a clean surface to carry out the functional test
correctly. Shake the control solution well before use. Undo the
cap and squeeze two drops next to each other on the clean
surface without touching them. Use the second drop for the
measurement.
Note
Never apply the control solution straight from the bottle to a test strip. Reason: the remaining
solution in the bottle will be contaminated if the top of the bottle comes into contact with the
test strip.
4
Apply drop to the test strip
Hold the input gap (at the tip of the test strip) to the drop of control solution until the gap
is completely lled. When the gap is suciently filled with solution, the device performs a
measurement. The device counts down for approx. five seconds. The measurement is then
shown on the display.
5
Evaluate result of functional test
Check whether the result is within the specified range of results for the control solution. This
range of results is printed on the test strip box or the test strip packaging or on the information
sheet included.
Expected results
At room temperature, the measured values from the test using the control solution should be within
the range printed on the test strip box in approx. 95% of all tests.
Warning
The result range printed on the test strip box applies only to the control solution. This is not a
recommended value for your blood glucose level.
30 Beurer GL50 evo
If measurements are outside the specified range, check the following possible causes:
Cause Action
The first drop of control solution was not
disposed of.
The tip of the bottle was not cleaned correctly.
The bottle was not shaken well enough.
Rectify the cause and repeat the test.
Control solution or test strip is contaminated.
Repeat the test using a new bottle of control
solution and/or new test strip.
The control solution, test strip or blood glucose
monitor is too warm or too cold.
Bring the control solution, test strip and monitor
to room temperature (+20°C to +26°C) and
repeat the test. The check performed at room
temperature is used as a general functional
check. The operating range specified in
the technical specifications is valid without
restriction.
The test strips and control solution were kept
at a temperature and humidity outside the
specified range.
Repeat the test using new, correctly stored
accessories (test strips and control solution).
Damaged test strips. Possible causes include
Test strips were exposed to fresh air for too
long.
The test strip box was not closed completely.
Repeat the test using a new test strip and/or a
new box of correctly stored test strips.
Test strip or control solution has expired.
Repeat the test using a new bottle of control
solution and/or new test strip from a new box.
Functional test using control solution was
performed incorrectly.
Repeat the test and follow the instructions.
Problem with the blood glucose monitor Contact Customer Services.
Warning
If you repeatedly obtain measured values outside the specified range when using control solution,
discontinue using the system to measure your blood glucose level. Contact Customer Services.
6 M easureMent MeMory
For each measurement, your blood glucose value is automatically saved with the date and time unless
” was activated for a blood glucose measurement using control solution.
The memory can store a maximum of 480 measured values. If the memory is full, the oldest value is
replaced by the most recent value. You can call up every individual measured blood glucose value.
You can also calculate and display the average blood glucose value for the last 7, 14, 30 and 90 days.
Beurer GL50 evo 31
Note
If you have already saved measured values and you reset the date, the average values are calculated
as from the new period.
“No saved values” indicates an empty memory for measured values. Press the ON/OFF button to
switch o the device.
6.1 Displaying individual values
The most recent measured value is displayed rst, and the oldest last. The date and time are also
displayed on the monitor at the same time.
1
Switch on the blood glucose monitor by briefly pressing and holding the scroll wheel.
2
Rotate the scroll wheel until the arrow is pointing towards “All values”, then press the scroll
wheel to confirm.
• All values
• Average
• Settings
Back
MEMORY
MEMORY
3
You can now rotate the scroll wheel to navigate between the saved individual values.
4
You can cancel the process at any time. To do so, press and hold the scroll wheel or wait
until the device switches itself o automatically after 2 minutes.
6.2 Displaying average blood glucose values for marked values
You can display the average measured blood glucose value of the individual markings from the last
7, 14, 30 and 90 days.
1Switch on the blood glucose monitor by briefly pressing and holding the scroll wheel.
2Rotate the scroll wheel until the arrow is pointing towardsAverage”, then press the scroll
wheel to confirm.
3 Use the scroll wheel to select the desired marking ( or O, , ) for which you want to
display the average values.
Ø 7 days 234 mg/dL O
Ø 14 days 156 mg/dL
Ø 30 days 107 mg/dL
Ø 90 days 98 mg/dL
Ø 7 days 13,0 mmol/L O
Ø 14 days 8,7 mmol/L
Ø 30 days 5,9 mmol/L
Ø 90 days 5,4 mmol/L
4
Press the scroll wheel to display the respective quantity of
measured values in the markings.
7 days - 4 values
14 days - 10 values
30 days - 109 values
90 days - 407 values
32 Beurer GL50 evo
6.3 Evaluating measured values on a PC
The GL50 evo monitor features an integrated plug-in USB stick. The blood glucose evaluation software
GlucoMemory is installed on the USB stick (for position of the USB connection, see page 11). The
GL50 evo is compatible with Diabass and SiDiary.
The blood glucose evaluation software GlucoMemory is pre-installed on the monitor's USB stick.
You can also install the software locally on a PC. This software enables you to evaluate your saved
measured values, add insulin doses with manual entries and print or export your results as a PDF or
CSV file. The software helps you and your GP to better monitor your blood glucose level.
For more information, please read software manual for the GlucoMemory software, including all the
necessary information and a detailed description of how to use the software (in English and German
language).
Note
An eective evaluation is only possible if you have set the date and time correctly (seeSetting the
date and time” on page 15).
Measurements cannot be taken while the USB stick is connected to a PC.
The measurements remain saved on the blood glucose monitor when the USB stick is removed from
the PC.
It is not possible to save software entries on the USB stick. Values are read only. If you want to make
entries, please first install the software on your PC.
Evaluating measured values on the PC
1The measuring device must be switched on. Insert
the monitor's USB connector into a free USB port on
your PC. In the event that the measuring device is not
recognised, please try using another active USB port.
2 ” is shown on the display of the monitor. Saved
data can now be viewed on your PC.
3See the information on evaluating values in the software manual.
Alternatively, you can transfer the measured values to your smartphone/tablet using the optional
Bluetooth® cap. The Beurer HealthManager app and software is available as a free download. For
further information, please read the instructions for use included with the Bluetooth
® adapter.
7 s , torinG MaintaininG and disinfectinG the device
Storing
Keep the Beurer GL50 evo blood glucose monitor in the case supplied after each measurement and
do not expose it to direct sunlight.
Beurer GL50 evo 33
Note
Do not store the device, test strips and control solution in your vehicle, in the bathroom or in a cooling
appliance.
Retain these instructions for use.
Only clean the device when it is switched o.
The device needs to be in the location in which the measurement will be taken for at least 30 minutes
before use so that the device can adapt to the ambient temperature. Please ensure that the operating
range is between +10°C and +40°C. Accurate measurement results cannot be guaranteed if a
measurement is taken outside of this temperature range.
7.1 Maintenance
Clean the surface of the device using a soft, slightly damp cloth (water or a mild cleaning solution).
Dry the device using a lint-free cloth.
Make sure that moisture does not enter the test strip insertion slot. Do not spray cleaning products
directly on the device. Do not submerge the device in water or any other fluids and make sure that
no fluids can get into the device.
7.2 Disinfection
Please comply with the generally applicable guidelines on disinfection when using the device on
dierent persons. Do not submerge the device in disinfection solutions or any other fluids and make
sure that no fluids can enter the device.
The cap on the integrated lancing device can be disinfected with 70 75% alcohol. Disinfect the cap
at least 1 x a week and submerge the cap in alcohol for approx. 10 minutes. Allow the cap to air dry.
Note
The monitor is made of precision components. The accuracy of the measurements and service life
of the device depend on its careful handling:
Protect the device from impacts and do not drop it.
Protect the device from damaging factors such as moisture, dirt, dust, blood, control solution or
water, marked temperature fluctuations, direct sunlight and extreme cold.
Do not use the device in the vicinity of strong electromagnetic elds, radio systems or mobile
telephones.
If the device is used in a dry environment, in particular near synthetic materials (clothes containing
synthetic fibres and carpets, for example), the damaging static charges which occur may cause
erroneous results.
Do not use the device near sources of strong electromagnetic radiation, as this may aect normal
operation.
It is a good idea to carry out an assessment of the electromagnetic environment before using the
device commercially.
34 Beurer GL50 evo
8 what if there are probleMs ?
Display messages on batteries and blood glucose measurement
No. Cause Solution
Empty
battery
Batteries discharged.
Charge the blood glucose monitor via the USB
interface.
High
temper-
ature
Temperature of the measuring
environment, blood glucose monitor
or test strip above the permitted
range.
Repeat the test using a new test strip as soon
as the measuring environment, monitor and
test strip have reached room temperature
(+20°C to +26°C). The check performed
at room temperature is used as a general
functional check. The operating range specified
in the technical specifications is valid without
restriction.
Low
temper-
ature
Temperature of the measuring
environment, blood glucose monitor
or test strip below the permitted
range.
Repeat the test using a new test strip as soon
as the measuring environment, monitor and
test strip have reached room temperature
(+20°C to +26°C). The check performed
at room temperature is used as a general
functional check. The operating range specified
in the technical specifications is valid without
restriction.
Strip
error
Used or contaminated test strip
inserted.
Insert an unused test strip that has not expired.
Repeat the blood glucose measurement.
Error 1 System error. Contact Customer Services.
Error 2
Insucient amount of blood on the
test strip.
Repeat the measurement using a new test
strip.
Error 5 System error. Contact Customer Services.
Unknown error messages. Contact Customer Services.
Beurer GL50 evo 35
Problem: device does not switch on
Cause Solution
ON/OFF switch set to o. Move the ON/OFF switch to the right.
Battery dead. Charge the blood glucose monitor via the USB
port on a computer.
Test strip inserted incorrectly or not completely.
Firmly insert the test strip into the slot on the
device with the contacts first. Make sure that
the front of the test strip is facing you (see Test
strips” on page 13).
Device faulty. Contact Customer Services.
Problem: the test does not start after inserting the test strip into the device and applying blood
Cause Solution
Insucient blood or test strip not filled correctly.
Repeat test using a new test strip and a larger
drop of blood.
Faulty test strip. Repeat the test using a new test strip.
Blood was applied while the device was
switched o.
Repeat the test using a new test strip.
The basic settings of the device have been
changed and these changes were not completed
(see “4.2 Making and changing basic settings”
on page 15).
Remove test strips and press and hold the scroll
wheel until the device is switched o.
Device faulty. Contact Customer Services.
36 Beurer GL50 evo
9 t echnical specifications
Dimensions (L x W x H) 123 x 29 x 16 mm
Weight 36 g
Power supply Li-ion rechargeable battery, 160 mAh
Battery life
150 measurements with fully charged rechargeable
battery
Measured value memory 480 measured values with date/time
Data retained when batteries are changed
Average values For 7, 14, 30, 90 days
Automatic switch-o 2 minutes after last actuation
Storage/transport temperature Temperature +2°C – +30°C
Relative humidity: < 90%
Operating ranges Temperature +10°C – +40°C
Relative humidity: < 90% non-condensing
Measurement range, glucose Glucose: 20 630 mg/dL (1.1 - 35.0 mmol/L)
Blood sample Capillary whole blood, venous whole blood
Required volume of blood 0.6 microlitres
Blood glucose measurement
duration Approx. 5 seconds
Calibration Plasma
Testing method Amperometric bio sensor
Usage Suitable for self-testing
System function test Each time device is switched on
The serial number is located on the device or in the battery compartment.
EMC
This device complies with the European standard EN 61326 and is subject to specific precautions
with regard to electromagnetic compatibility. Please note that portable and mobile HF communication
systems may interfere with this device. For more details, please contact our Customer Services at
the address indicated.
Test strip functionality
Test strips enable a quantitative measurement of the glucose level in fresh whole blood (capillary or
venous). When the gap for taking blood comes into contact with a drop of blood, it is automatically
filled by simple capillary action. The blood is sucked into the absorbing gap on the test strip and the
monitor measures the glucose level in the blood.
The test is based on the measurement of an electric current that is generated by the chemical reaction
of the glucose with the enzyme glucose dehydrogenase (Aspergillus oryzae) on the strip.
Beurer GL50 evo 37
During the reaction, a mediator transports electrons through the electrode surface and so generates
a current.
The monitor analyses this current. The current ow is proportional to the glucose content in the blood
sample. The results are shown on the blood glucose monitor display. Only a small amount of blood
is required (0.6 microlitres) and measurement takes approx. 5 seconds. The test strip detects blood
glucose values from 20 to 630 mg/dL.
Chemical components of the test strip sensor
FAD glucose dehydrogenase 6%
Potassium ferricyanide 56%
Non-reactive components 38%
Control solution functionality
The control solution contains axed amount of glucose that reacts with the test strip. A test with
control solution is similar to a blood test. However, control solution is used instead of blood. The
measurement using control solution must be within the result range. This range is printed on each
test strip box.
Chemical composition of the control solution
The control solution is a red solution with the following D-glucose level (in percentage shares).
Substances LEVEL 3 control solution LEVEL 4 control solution
D-glucose 0.14% 0.37%
Non-reactive components 99.86% 99.63%
Standards
The Beurer GL50 evo blood glucose monitor complies with the European standards:
IVD (98/79/EEC) and MDD (93/42/EEC).
10 coMparison of Measured values with laboratory values
Precision
Three lots of the GL50 evo blood glucose test strips have been tested to assess the precision of the
GL50 evo blood glucose monitoring system. This includes a repeat assessment using venous blood
and a laboratory precision assessment using the control material. The blood glucose content of the
venous blood samples ranges from 42.7 to 418.0 mg/dL (2.37 to 23.20 mmol/L) and control material
from three concentrations is used.
38 Beurer GL50 evo
Results of the repeat precision measurements
Sample
Venous blood Overall average
value
Pooled standard
deviation
Pooled coecient
of variation (%)
mg/dL mmol/L mg/dL mmol/L mg/dL mmol/L
1 42.7 2.4 36.0 2.0 2.0 0.1 5.6
2 62.0 3.4 59.2 3.3 3.5 0.2 5.9
3 120.5 6.7 127.1 7.1 4.1 0.2 3.2
4 201.0 11.2 213.8 11.9 6.7 0.4 3.1
5 316.5 17.6 329.9 18.3 10.1 0.6 3.1
6 418.0 23.2 433.5 24.1 14.5 0.8 3.3
Results of the intermediate precision measurement
Sample
Control material Overall average
value
Pooled standard
deviation
Pooled coecient
of variation (%)
mg/dL mmol/L mg/dL
mmol/L
mg/dL mmol/L
1 70.0 3.9 71.3 4.0 1.0 0.1 1.4
2 135.6 7.5 136.3 7.6 1.4 0.1 1.1
3 351.5 19.5 350.8 19.5 2.8 0.2 0.8
System accuracy
The GL50 evo blood glucose monitor in comparison to the YSI.
Three lots of GL50 evo blood glucose test strips have been tested to assess the system accuracy
of the GL50 evo blood glucose monitoring system and to compare it with the reference method, in
which capillary whole blood concentrations of 36.0 mg/dL (2.0 mmol/L) to 442.5 mg/dL (24.6 mmol/L)
have been used.
Results of the system accuracy for glucose concentrations <100 mg/dL (<5.55 mmol/L)
Within ±5 mg/dL
(Within ±0.28 mmol/L)
Within ±10mg/dL
(Within ±0.56 mmol/L)
Within ±15mg/dL
(Within ±0.83 mmol/L)
101/168 (60.12%) 161/168 (95.83%) 166/168 (98.81%)
Results of the system accuracy for glucose concentrations ≥100 mg/dL (≥5.55 mmol/L)
Within ±5% Within ±10% Within ±15%
182/432 (42.13%) 358/432 (82.87%) 426/432 (98.61%)
Beurer GL50 evo 39
Results of the system accuracy for combined glucose concentrations between 36.0 mg/dL
(2.0 mmol/L) and 442.5 mg/dL (24.6 mmol/L).
Within ±15 mg/dL or ±15%
(within ±0.83 mmol/L or ±15%)
592/600 (98.67%)
In comparison to the YSI, the GL50 evo met the EN ISO 15197:2015 standard, whereby 95%
of the blood glucose values measured have to fall within the following zones: either ±15 mg/dL
(± 0.83 mmol/L) of the measured average values when using the reference measuring procedure for
blood glucose concentrations <100 mg/dL (<5.55 mmol/L) or ±15% for blood glucose concentrations
of ≥100 mg/dL (≥5.55 mmol/L). 99% of the individual measured blood glucose values must fall within
zones A and B of the Consensus Error Grid (CEG) for diabetes type 1.
Performance evaluation by the user
A study to assess the glucose values of blood samples of capillary blood from the fingertips, which
were obtained from 103 individuals that had no special training, produced the following results:
96.7% within ±15 mg/dL 0.83 mmol/L) and 95.9% within ±15% of the values obtained in the medical
laboratory with glucose concentrations of at least 100 mg/dL (5.55 mmol/L).
You will find further details and information regarding blood glucose results and various technologies
in generally relevant specialist medical literature.
11 u saGe liMits for specialist personnel froM the
healthcare sector
1. If the patient exhibits the following symptoms, it may be the case that no correct values can be
obtained:
• Acute dehydration
• Acute hypotension (low blood pressure)
• Shock
• Hyperosmolar hyperglycaemic condition (with or without ketosis)
2. Lipaemic samples: cholesterol levels up to 500 mg/dL and triglyceride levels up to (13 mmol/L)
1000 mg/dL (11.4 mmol/L) do not influence the results. Severely lipaemic blood samples were not
tested with the Beurer GL50 evo blood glucose monitoring system; therefore, using the device with
these samples is not recommended.
3. In the case of severely ill patients, blood glucose monitors for home use should not be used.
4. The eect of interfering substances on the measurements depends on the concentration in the
blood. The maximum concentrations of certain substances listed below do not significantly
influence the measurements.
40 Beurer GL50 evo
Concentration of
tested substances
50-100 mg/dL
(2.8-5.6 mmol/L)
250-350 mg/dL
(13.9-19.4 mmol/L)
Acetaminophen 7 mg/dL (0.46 mmol/L) 6.6 mg/dL
(0.37 mmol/L)
4.5%
Ascorbic acid 4 mg/dL (0.23 mmol/L) 3.3 mg/dL
(0.18 mmol/L)
5.1%
Bilirubin 3.3 mg/dL (0.06 mmol/L) 0.1 mg/dL
(0.01 mmol/L)
-1.4%
Cholesterol 400 mg/dL (10.34 mmol/L) -6.8 mg/dL
(-0.38 mmol/L)
-6.2%
Creatinine 30 mg/dL (2.65 mmol/L) 0.0 mg/dL
(0.00 mmol/L)
-0.1%
Dopamine 2.2 mg/dL (0.14 mmol/L) 5.0 mg/dL
(0.28 mmol/L)
1.0%
EDTA 5.0 mg/dL (0.17 mmol/L) -2.0 mg/dL
(-0.11 mmol/L)
-2.4%
Ephedrine 40 mg/dL (2.42 mmol/L) -3.9 mg/dL
(-0.22 mmol/L)
2.4%
Galactose 20 mg/dL (1.11 mmol/L) -3.1 mg/dL
(-0.17 mmol/L)
0.5%
Gentisic acid 7 mg/dL (0.45 mmol/L) 7.2 mg/dL
(0.40 mmol/L)
2.9%
Glutathione 1 mg/dL (0.03 mmol/L) -2.6 mg/dL
(-0.14 mmol/L)
-3.7%
Haemoglobin 300 mg/dL (0.05 mmol/L) -3.1 mg/dL
(-0.17 mmol/L)
-2.6%
Heparin 2.1 mg/dL (0.0018 mmol/L) -3.0mg/dL
(-0.17 mmol/L)
-1.3%
Ibuprofen 50 mg/dL (2.43 mmol/L) -2.6 mg/dL
(-0.15 mmol/L)
-1.9%
Icodextrin 1094 mg/dL
(0.64~0.78 mmol/L)
-4.17 mg/dL
(-0.23 mmol/L)
-2.9%
Blood glucose
valueInfluence
Beurer GL50 evo 41
Concentration of
tested substances
50-100 mg/dL
(2.8-5.6 mmol/L)
250-350 mg/dL
(13.9-19.4 mmol/L)
L-Dopa 2 mg/dL (0.10 mmol/L) 9.3 mg/dL
(0.52 mmol/L)
7.9%
Maltose 278 mg/dL (7.72 mmol/L) -1.53 mg/dL
(-0.09 mmol/L)
-2.6%
Methyldopa 4 mg/dL (0.19 mmol/L) 7.3 mg/dL
(0.41 mmol/L)
0.9%
Pralidoxime iodide 5 mg/dL (0.14 mmol/L) 1.7 mg/dL
(0.09 mmol/L)
-0.1%
Sodium salicylate 40 mg/dL (2.50 mmol/L) -3.1 mg/dL
(-0.17 mmol/L)
-0.6%
Salicylic acid 60 mg/dL (4.34 mmol/L) -0.1 mg/dL
(-0.01 mmol/L)
7.6%
Tolbutamide 100 mg/dL (3.70 mmol/L) 0.5 mg/dL
(0.03 mmol/L)
-0.8%
Tolazamide 2.5 mg/dL (0.08 mmol/L) -2.3 mg/dL
(-0.13 mmol/L)
1.8%
Triglyceride 800 mg/dL (9.37 mmol/L) -7.50 mg/dL
(-0.42 mmol/L)
-4.0%
Uric acid 16.5 mg/dL (0.98 mmol/L) 6.6 mg/dL
(0.37 mmol/L)
1.8%
Xylose 9.5 mg/dL (0.63 mmol/L) 5.6 mg/dL
(0.31 mmol/L)
6.6%
12 w / arranty service
Beurer GmbH, flinger Straße 218, 89077 Ulm, Germany (hereinafter referred to as Beurer”) provides
a warranty for this product, subject to the requirements below and to the extent described as follows.
The warranty conditions below shall not affect the sellers statutory warranty obligations which
ensue from the sales agreement with the buyer.
The warranty shall apply without prejudice to any mandatory statutory provisions on liability.
Blood glucose
valueInfluence


Specyfikacje produktu

Marka: Beurer
Kategoria: miernik glukozy we krwi
Model: GL 50 evo
Kolor produktu: Czarny
Technologia baterii: Alkaliczny
Liczba baterii: 2
Połączenie z pc (personal computer): Tak
Funkcja pamięci: Tak
Czas mierzenia: 5 s
Liczba zapisów pamięci: 480
Jednostki pomiaru glukozy: mg /dL
Rozmiar próbki krwi: 0.6 µl

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Beurer GL 50 evo, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą