USER MANUAL
ECO209 - pH Monitor
ISTECO209 09/14
WARRANTY
This instrument is warranted from all defects in materials and
manufacturing for a period of 2 years from the date of purchase.
Electrodes and probes are warranted for a period of 6 months.
If during this period, the repair or replacement of parts is required,
where the damage is not due to negligence or erroneous operation by
the user, please return the parts to either dealer or our office and the
repair will be effected free of charge.
Note: We reserve the right to modify the design, construction and
appearance of our products without advance notice.
SPECIFICATIONS
Range 0.0 to 14.0 pH
Resolution 0.1 pH
Accuracy (@25°C) ±0.2 pH
pH electrode AD1230B
Calibration Manual, 2-point, through trimmers
Environment 0 to 50 ºC, 95% RH max
Power supply External, 12 Vdc
Dimensions 120 x 134 x 54 mm
Weight approx. 250 g
ACCESSORIES
AD1230B pH electrode with plastic body, double junction BNC
connector and 2 m cable
AD70004P pH 4.01 buffer solution, 20 ml sachet (25 pcs.)
AD70007P pH 7.01 buffer solution, 20 ml sachet (25 pcs.)
AD70010P pH 10.01 buffer solution, 20 ml sachet (25 pcs.)
AD70000P Rinse solution, 20 ml sachet (25 pcs.)
GENERAL OPERATIONS
The unit is supplied complete with AD1230B pH electrode.
Connect the electrode to the BNC connector on the bottom of
the meter. Do not be alarmed if white crystals appear around
the electrode protective cap. They dissolve by simply rinsing
with water.
If the electrode is dry, soak the tip in electrode rinse solution
for one hour to reactivate it.
Connect the supplied 12 Vdc adapter to the power supply
connector and to the mains.
Make sure the unit has been calibrated for pH before using it.
The meter is supplied with a differential input and a stainless
steel probe to prevent potential grounding problems.
For taking measurements remove the protective cap from the
pH electrode and immerse the tips (4 cm) of both pH electrode
and grounding probe into the solution to be measured, then
shake briefly.
The display will show the pH value. Wait a few seconds for
the reading to stabilize and the meter will start continuous
monitoring.
To turn the unit off simply unplug the 12 Vdc adapter from the
mains.
CALIBRATION PROCEDURE
A new pH meter must always be calibrated before use.
The ECO209 pH-Monitor can be calibrated manually over 2 points,
at pH 7 and pH 4 (or pH 10).
If you are going to measure acidic samples (pH value between
0 and 7), it is suggested to calibrate with pH 7.01 and pH 4.01
buffer solutions.
If your samples are alkaline (pH value between 7 and 14), use
pH 7.01 and pH 10.01 buffer solutions.
Immerse the tips of electrode and grounding
probe in a new pH 7 calibration solution and
allow the reading to stabilize.
Using a small screwdriver adjust the pH 7
calibration trimmer to read pH "7.0".
Rinse the electrode in the rinse solution.
Immerse the tips of electrode and grounding
probe in a new pH 4 (or pH 10) solution and
wait for the reading to stabilize.
Adjust the pH 4 calibration trimmer to read
pH "4.0" (or pH "10.0").
The pH calibration procedure is now complete and the meter
is ready for use.
CALIBRAT ION 
SOLUTION
pH ELECTRODE MAINTENANCE
The electrode should always be kept moist and wet. If the
electrode dries out, soak it in rinse solution for at least 1 hour.
When not in use, rinse the electrode with water and store it
with a few drops of pH 7 or rinse solution in the protective
cap.
Always replace the protective cap after use.


Specyfikacje produktu

Marka: Adwa
Kategoria: Sprzęt pomiarowy
Model: ECO209

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Adwa ECO209, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Sprzęt pomiarowy Adwa

Adwa

Adwa AD1040 Instrukcja

5 Października 2024
Adwa

Adwa ECO406 Instrukcja

5 Października 2024
Adwa

Adwa AD31 Instrukcja

5 Października 2024
Adwa

Adwa AD100 Instrukcja

5 Października 2024
Adwa

Adwa AD201 Instrukcja

5 Października 2024
Adwa

Adwa AD203 Instrukcja

5 Października 2024
Adwa

Adwa AD32 Instrukcja

21 Września 2024
Adwa

Adwa AD14 Instrukcja

21 Września 2024
Adwa

Adwa AD3000 Instrukcja

20 Września 2024
Adwa

Adwa ECO209 Instrukcja

20 Września 2024

Instrukcje Sprzęt pomiarowy

Najnowsze instrukcje dla Sprzęt pomiarowy

Stamos

Stamos S-LS-28 Instrukcja

15 Października 2024
Stamos

Stamos S-LS-86 Instrukcja

15 Października 2024
Stamos

Stamos S-LS-61 Instrukcja

15 Października 2024
Stamos

Stamos S-LS-81 Instrukcja

15 Października 2024
Stamos

Stamos S-LS-84 Instrukcja

15 Października 2024
Stamos

Stamos S-LS-74 Instrukcja

15 Października 2024
Stamos

Stamos S-LS-76 Instrukcja

15 Października 2024
Stamos

Stamos S-LS-42 Instrukcja

15 Października 2024
Stamos

Stamos S-LS-75 Instrukcja

15 Października 2024
Stamos

Stamos S-LS-35 Instrukcja

15 Października 2024