Owner’s Manual
Isolator® Series 115/230V
Medical-Grade™
Isolation Transformers
with Step-Up and Step-Down Capabilities
IS300HGDV
(Series Number AG-8798)
IS600HGDV
(Series Number AG-8799)
IS1000HGDV
(Series Number AG-879A)
IS1800HGDV
(Series Number AG-879B)
TrippLite.Eaton.com/support
Copyright © 2023 Tripp Lite. All rights reserved. All trademarks are the sole property of their respective owners.
Table of Contents
1. Overview 2
2. Important Safety Instructions 2
3. Installation 3
4. Optional Connections 5
5. Troubleshooting 5
6. Maintenance 6
7. Warranty 6
Español 7
Français 13
Русский 19
Deutsch 25
2
1. Overview
2. Important Safety Instructions
Statement of Intended Use
Tripp Lite 115/230V ANSI/AAMI/CSA 60601-1 / IEC 60601-1 3.1 ed. Listed Medical-Grade Isolation Transformers offer line isolation, step-up
and step-down capabilities and continuous noise ltering. An internal low-impedance isolation transformer component offers 100% isolation
from the input AC line. Full ANSI/AAMI/CSA 60601-1 / IEC 60601-1 3.1 ed. medical-grade listing makes it ideal for protection of medical
electronic equipment in patient-care vicinities. Faraday shield and oating AC output reduces the cumulative leakage current of the Isolator
and connected equipment to levels less than 100 microamps. Transformer ltering provides continuous common-mode noise rejection with
no wearable parts. All units are equipped with a potential equalization bolt (POAG) and ground lug for connection to a medical potential
equalization installed in the room. Rugged all-metal unit housing; ships ready for placement in any industrial, medical, ofce or network
environment. Step-up and step-down transformer (115/230V 50/60Hz) with universal IEC-320-C14 or IEC-320-C20 inputs and
IEC-320-C13 outputs allows for worldwide use.
The above features also make these products ideal for audio-video isolation, as well as other similar applications.
Transformer Location Warnings
Install the transformer indoors, away from excess moisture or heat, conductive contaminants, dust or direct sunlight.
For best performance, keep the indoor temperature at 77°F (25°C). The operational range is between 32° and 104°F (0° and 40°C).
Transformer Connection Warnings
Connect the transformer directly to a properly grounded AC power outlet.
Do not modify the transformer’s plug, and do not use an adapter that would eliminate the transformer’s ground connection.
To remove AC voltage from the transformer, pull the plug from the socket outlet.
Remove mains supply before installing, uninstalling, fuse replacement or system changes.
Observe accident prevention and safety regulations that apply specially to the application.
Before start-up check whether the rated voltage of the equipment matches the output voltage selection on the transformer.
Do not use this equipment within oxygen-enriched atmospheres or within 1 ft. (0.3 m) at a point from which an oxygen-enriched atmosphere
is intentionally ventilated.
Equipment Connection Warnings
Use of this equipment in life support applications where failure of this equipment can reasonably be expected to cause the failure of the life
support equipment or to signicantly affect its safety or effectiveness is not recommended.
To reduce the risk of re, connect only to a circuit that has branch circuit over current protection with an ampere rating in accordance with
the National Electrical Code® (NEC®), ANSI/NFPA 70 or your local electrical code. In Europe, the circuit breaker must meet the IEC/EN
60934 standard and have a contact air gap of at least 3 mm.
3
1
2
3
1
2
3
Use a small, at blade screwdriver to open the
access cover and fuse holder for the fuses and
to set the output voltage.
Use a small, at blade screwdriver to open the
access cover and fuse holder for the fuses and
to set the output voltage.
3. Installation
Before connecting equipment to the isolation transformer, verify the local mains voltage matches the input voltage
selected on the transformer. The output voltage must match the voltage compatible with the equipment load.
The appliance inlet of the isolation transformer (300VA, 600VA or 1000VA) is designed so power cannot be applied while
the fuse access door is opened. If this protective feature is damaged, have the appliance inlet replaced by a qualied
service person. Do not defeat this protective feature or attempt to perform inspection or replacement of fuses while the
isolation transformer is ON and plugged into mains power.
Install in accordance with applicable electrical codes, such as National Electrical Code and/or NFPA 99 Article 517. Use only with an approved
plug/cord set according to the equipment electrical ratings being used. The transformer must be installed by a qualied technician only.
Compliance with ANSI/AAMI/CSA 60601-1 / IEC 60601-1 3.1 ed. must be evaluated after the nal installation.
3.1 Select input and output voltage on the isolation transformer based on the local mains
voltage available and the voltage rating of the equipment to be connected.
3.1.1 For IS300HGDV and IS600HGDV:
Input voltage selection is made via the inlet fuse holder (Figure 1). The voltage selected will display above the power switch A. You will
also have to insert the correct fuse value depending on the transformer model and input voltage. For correct fuse value, see the table
in section 3.1.5. Output voltage selection is made with the selector switch B.
If using as a step-up or step-down transformer, the output voltage must match the connected equipment load requirements.Note:
Figure 1
Figure 2
A
A
B
B
Output Voltage
Selector Switch
Output
Voltage
Selector
Switch
3.1.2 For IS1000HGDV:
Input voltage selection is made via the inlet fuse holder (Figure 2). The voltage selected will display above the power switch A. You
will also have to insert the correct fuse value depending on input voltage. For correct fuse value, see the table in section 3.1.5. Output
voltage selection is made with the selector switch B.
Note: If using as a step-up or step-down transformer, the output voltage must match the connected equipment load requirements.


Specyfikacje produktu

Marka: Tripp Lite
Kategoria: nieskategoryzowany
Model: Isolator IS1800HGDV
Częstotliwość wejściowa AC: 50 - 60 Hz
Napięcie wejściowe AC: 115 - 230 V
Wysokość produktu: 140 mm
Szerokość produktu: 300 mm
Głębokość produktu: 350 mm
Waga produktu: 23470 g
Szerokość opakowania: 445 mm
Wysokość opakowania: 279.9 mm
Głębokość opakowania: 500.1 mm
Certyfikaty: ANSI/AAMI ES60601-1:2005+A2+A1, IEC60601-1 3RD EDITION +A1, cAN/CSA-C22.2 NO. 60601-1:14, CE, NOM
Kraj pochodzenia: Tajwan
Zakres temperatur (eksploatacja): 0 - 40 °C
Zakres wilgotności względnej: 5 - 95 %
Waga wraz z opakowaniem: 23470 g
Wysokość skrzyni wzorcowej (zewnętrznej): 279.9 mm
Długość skrzyni głównej (zewnętrznej): 500.1 mm
Waga brutto obudowy głównej (zewnętrznej): 23230 g
Szerokość skrzyni zbiorczej (zewnętrznej): 445 mm
Zakres temperatur (przechowywanie): -15 - 50 °C
Dopuszczalna wysokość podczas eksploatacji (n.p.m.): 0 - 3000 m
Kolor obudowy: Biały
Pojemność: 1800 VA
Produkty w skrzyni głównej (zewnętrznej): 1 szt.
Ilość gniazd sieciowych: 8 x gniazdo sieciowe
Typy wyjść AC: C13
Zabezpieczenie przed zwarciem: Tak

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Tripp Lite Isolator IS1800HGDV, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje nieskategoryzowany Tripp Lite

Instrukcje nieskategoryzowany

Najnowsze instrukcje dla nieskategoryzowany

Meec Tools

Meec Tools 010-084 Instrukcja

6 Października 2024
Outwell

Outwell Scenic Road 250 Instrukcja

5 Października 2024
Impact

Impact QuikBalance Instrukcja

4 Października 2024
Impact

Impact CA-112 Instrukcja

4 Października 2024
Impact

Impact LS-BHA Instrukcja

4 Października 2024
Impact

Impact LKB-5C Instrukcja

4 Października 2024
Impact

Impact RST-001 Instrukcja

4 Października 2024