Instrukcja obsługi Sauter HMM-NP


Przeczytaj poniżej 📖 instrukcję obsługi w języku polskim dla Sauter HMM-NP (13 stron) w kategorii Sprzęt pomiarowy. Ta instrukcja była pomocna dla 5 osób i została oceniona przez 2 użytkowników na średnio 4.5 gwiazdek

Strona 1/13
Sauter GmbH
Ziegelei 1
D-72336 Balingen
e-mail: info@kern-sohn.com
Phone : +49-[0]7433- 9933-0
Fax: +49-[0]7433-9933-149
Internet: www.sauter.eu
Instruction manual mobile Leeb
hardness tester
SAUTER HMM
Version 2.0
04/2020
GB
HMM-BA-e-2020
PROFESSIONAL MEASURING
2 HMM-BA-e-2020
GB SAUTER HMM
V. 2.0 04/2020
Instruction manual mobile Leeb hardness tester
Thank you for purchasing the mobile digital Leeb impact hardness tester from
SAUTER. We hope that you will be very satised with the high quality of this device
and its extensive functionality. For any questions, wishes and suggestions we are at
your disposal.
Table of contents:
1
Before commissioning ................................................................................... 3
2
Introduction .................................................................................................... 3
2.1 Measuring principle ............................................................................................................. 3
2.2 Structure .............................................................................................................................. 4
2.3 Keys ..................................................................................................................................... 4
2.4 Display overview.................................................................................................................. 5
2.5 Technical data ...................................................................................................................... 5
3
Test preparation ............................................................................................. 6
3.1 Preparation of the test object .............................................................................................. 6
3.2 Sample thickness ................................................................................................................ 7
3.3 Test procedure ..................................................................................................................... 7
3.3 Test execution ..................................................................................................................... 8
3.4 Browse ................................................................................................................................. 8
3.5 Print test report .................................................................................................................... 9
4
Settings ........................................................................................................... 9
4.1 Material code........................................................................................................................ 9
4.2 Impact direction ................................................................................................................... 9
4.3 Scale ..................................................................................................................................... 9
5
Service and maintenance ............................................................................ 10
5.1 Periodic maintenance ........................................................................................................ 10
6
Calibration ..................................................................................................... 10
7
Annexes ........................................................................................................ 12
7.1 A-1 Conversion range ........................................................................................................ 12
7.2 A-2 Material Code .............................................................................................................. 12
7.3 A-3 Scope of Delivery ........................................................................................................ 13
HMM-BA-e-2020 3
1 Before commissioning
Before putting the device into operation, check the delivery for any transport damage
to the packaging, the plastic case and the device itself. Should this be the case,
SAUTER must be contacted immediately.
2 Introduction
2.1 Measuring principle
The HMM hardness tester is physically a rather simple dynamic hardness tester: A
rebound sensor with a hard metal tip is driven against the surface of the test object
with spring force. The surface can be damaged if the rebound body hits the surface,
which is ultimately due to a loss of kinetic energy.
This energy loss is calculated from the speed measurements when the rebound sensor
is at a certain distance from the surface for the rebound and trigger phases of the test.
The xed magnet in the rebound body generates an induced voltage in the simple wire
coil of the rebound body.
The voltage of the signal is proportional to the speed of the rebound sensor. The signal
processing by the electronics ensures that the hardness value can be read on the
display and stored.


Specyfikacje produktu

Marka: Sauter
Kategoria: Sprzęt pomiarowy
Model: HMM-NP

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Sauter HMM-NP, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Sprzęt pomiarowy Sauter

Sauter

Sauter FA 50 Instrukcja

5 Stycznia 2025
Sauter

Sauter FC 100 Instrukcja

5 Stycznia 2025
Sauter

Sauter TVP-L Instrukcja

5 Stycznia 2025
Sauter

Sauter FS 4 Instrukcja

5 Stycznia 2025
Sauter

Sauter 283-602 Instrukcja

5 Stycznia 2025
Sauter

Sauter FL 20K Instrukcja

5 Stycznia 2025
Sauter

Sauter DA 1-4 Instrukcja

5 Stycznia 2025
Sauter

Sauter 281-601 Instrukcja

5 Stycznia 2025
Sauter

Sauter TPE-N Instrukcja

5 Stycznia 2025

Instrukcje Sprzęt pomiarowy

Najnowsze instrukcje dla Sprzęt pomiarowy