1. Technical data
KERN -2 -1-A -3 TGC 150 TGC 500 TGC 1K
Item no./ Type -2-A -1-A -3-A TTGC 150 TGC 500 TGC 1K
Readability (d) 0.1 g 1 g 0.01 g
Weighing range 000 g(max) 500 150 g g 1
Adjustment weight, not added (class) / 150 g 00 g (M3) 0 g (M3) 50 g (M2) 5 100
Operating temperature 0+ °....+ 40°C
Humidity of air 20% 8 - 0% air humidity (non condensing)
Batteries 2 x CR2032
2. Scope of delivery
Pocket scale
Balance cover
Batteries (2 x CR2032)
Operating manual
3. Keyboard overview
Switch-over weighing unit
Background illumination of the display on/off
Taring / Setting to zero
Switch on/off
4. Operation
Weighing:
Place the balance on a firm, level surface.
Press ON/OFF button, the balance completes a self check. As soon as the weight display appears,
the balance is ready for weighing. If required, set to zero by using TARE button.
Place load on pan and read the weighing result.
If necessary, use the button to switch over into the available weighing units.UNIT
3 - -e- TGC BA 1810
Taring:
Deposit empty weighing container.
Press the TARE button, the zero display disappears.
Weigh the material, the net weight will be displayed.
When the balance is unloaded the saved taring value is displayed with negative sign.
To delete the stored tare value, remove load from weighing pan and press the TARE button.
5. Adjustment
Arrange adjustment weight (see chap. 1).
Info about adjustment weights can be found on the Internet at: http://www.kern-sohn.com
Ensure that there are no objects on the weighing pan.
In weighing mode keep the button pressed until UNIT <CAL> appears.
Press the ON/OFF button again and wait until the weight value of the required adjustment weight
will appear flashing.
Place adjustment weight. Wait until <PASS>, followed by the weight value of the placed
adjustment weight will be displayed. After that the adjustment will be finished successfully.
In case of a diverging display the adjustment has failed. Repeat adjustment process.
6. Important Notes
Do not overload balance.
The occurrence of electromagnetic fields may result in major display deviations. Then change
location.
Avoid high air humidity, vapours and dust.
Do not bring the balance directly in contact with liquids.
Do not leave permanent load on the weighing pan.
Never operate balance in explosive environment. The serial version is not
explosion protected.
Check the balance regularly with external test weights.
Warranty claims will be voided when the housing is opened.
7. Declaration of conformity
The current EC/EU Conformity declaration can be found online in:
www.kern-sohn.com/ce


Specyfikacje produktu

Marka: Kern
Kategoria: waga
Model: TGC 150-2

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Kern TGC 150-2, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje waga Kern

Kern

Kern FOB Instrukcja

10 Września 2024
Kern

Kern HCB 200K500 Instrukcja

10 Września 2024
Kern

Kern HCB 100K200 Instrukcja

10 Września 2024
Kern

Kern DE 6K1D Instrukcja

10 Września 2024
Kern

Kern DE 35K5DL Instrukcja

10 Września 2024
Kern

Kern HCB 50K20 Instrukcja

10 Września 2024
Kern

Kern MAP 80K1S Instrukcja

10 Września 2024
Kern

Kern PCB-A01 Instrukcja

10 Września 2024
Kern

Kern HCB 50K100 Instrukcja

10 Września 2024
Kern

Kern PCD-A04 Instrukcja

9 Września 2024

Instrukcje waga

Najnowsze instrukcje dla waga

Korona

Korona Sophie Instrukcja

15 Października 2024
Beurer

Beurer GS 209 Instrukcja

15 Października 2024
Tanita

Tanita UM-028F Instrukcja

15 Października 2024
Beurer

Beurer GS 51 XXL Instrukcja

14 Października 2024
Balance

Balance KH 807 Instrukcja

14 Października 2024
Tanita

Tanita UM-070 Instrukcja

13 Października 2024
Beper

Beper 40.801B Instrukcja

13 Października 2024
Taylor

Taylor 7548 Instrukcja

13 Października 2024
Tronic

Tronic KH 88 R Instrukcja

13 Października 2024
Jata

Jata P111 Instrukcja

13 Października 2024