Model 707
Loop Calibrator
Calibration Information
W�Warning
To prevent electrical shock, remove test leads and any input signals from the
Model 707 Loop Calibrator before opening the case.
WCaution
The Model 707 Loop Calibrator contains parts that can be damaged by static
discharge. Follow the standard practices for handling static sensitive
devices.
Introduction
The for the Model 707 Loop Calibration Information
Calibrator (hereafter referred to as the Loop
Calibrator) provides the information necessary to
verify the performance and adjust the calibration of
the Loop Calibrator.
This document includes the following:
Contacting Fluke for service
Specifications
Cleaning
Replacing the battery
Disassembly and reassembly
Recommended test equipment
Performance verification procedures
Calibration adjustments
User-replaceable parts
The complete operating instructions and warranty
statement for the Model 707 Loop Calibrator are
provided in the Instruction Sheet that came with it.
Contacting Fluke for Service
This Fluke product will be free from defects in
material and workmanship for three years from the
date of purchase.
To contact Fluke, call one of the following telephone
numbers:
Technical Support USA: 1-800 44 800 443 5853)- - -FLUKE (1 - -
Calibration/Repair USA: 1-888 99 888 993 5853)- - -FLUKE (1 - -
Canada: 1-800 36 800 363 5853)- - -FLUKE (1 - -
Europe: +31 402-675 200-
Japan: +81-3-6714 3114-
Singapore: +65-6799 5566-
China: +86-400 921 0835- -
Brazil: +55- - - 11 3530 8901
Anywhere in the world: +1-425 446 5500- -
Or, visit Fluke's website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To view, print, or download the latest manual supplement,
visit http://us.fluke.com/usen/support/manuals.
Address correspondence to:
Fluke Corporation Fluke Europe B.V.
P.O. Box 9090, P.O. Box 1186,
Everett, WA 98206-9090 5602 BD Eindhoven
U.S.A. The Netherland
Fluke 707
Calibration Information
2
Specifications
Product specifications are provided in Tables 1 and 2. These specifications are based on a 1-year calibration cycle
and apply from 18 ºC to 28 ºC (64 ºF to 82 ºF) unless stated otherwise. Accuracy specifications are given as
follows:
± ( [ % of reading ] + [ number of least significant digits ] ).
Table 1. General Specifications
Maximum voltage between any
terminal and earth ground or
between any two terminals
30 V
Display: 0 to 28 V with 0.001 V resolution, 0 to 24 mA with 0.001 mA resolution
Operating Temperature: -10 °C to 55 °C
Storage Temperature: -30 °C to 60 °C
Temperature Coefficient: ± 0.005 % of range per C to 18 d 28 C to 55 °C for -10 ° °C an ° °C
Relative Humidity: 95 % to 30 °C,
75 % to 40 °C,
45 % up to 50 °C,
35 % up to 55 °C
Operating Altitude: 3000 meters maximum
Power Requirements: Single 9 V battery (IEC 6LR61)
Battery Life (Typical): Source Mode: 18 hours; 12 mA into 500
Measure / Simulate Mode: 50 hours
Size (H x W x L):
With Holster:
50.80 mm x 69.85 mm x 142.87 mm (2.00 in x 2.75 in x 5.625 in)
54.61 mm x 76.20 mm x 158.75 (2.15 in x 3.00 in x 6.25 in)
Weight: 349 g (12.3 oz). With Holster and flex- stand: 601 g (21.2 oz)
Safety:
IEC 61010 1: Pollution degree 2-
IEC 61010 030: CAT O 30-2- V
Table 2. Accuracy Specifications
Function Range Resolution
Accuracy
(% of reading + counts)
DC Volts Input
+ 28 V (+30 V max) 0.001 V ± (0.015 % + 2 counts)
DC mA Input 20 mA (24 mA max) 0.001 mA ± (0.015 % + 2 counts)
DC mA Output 0 mA to 24 mA 0.001 mA ± (0.015 % + 2 counts)
Source mode: Compliance: ≥ Ω 1200 at 20 mA, 950 at 20 mA in Hart Mode
Simulate mode: External loop voltage requirements: 24 V nominal, 28 V maxi 2 V minimummum, 1
Loop Power 24 V
Percent display
- 25 % to 125 % (4 to 20 mA mode)
Input/Output protection Fuseless protection.
Loop Calibrator
Cleaning
3
Cleaning
W Warning
To prevent electrical shock, remove test
leads and input signals before cleaning.
To clean the case, wipe it with a cloth lightly
dampened with water and a mild detergent.
Do not use abrasives, solvents, or alcohol.
Replacing the Battery
WWarning
To avoid false readings, which could lead
to possible electric shock or personal
injury, replace the battery as soon as the
low battery symbol ( ) appears on the M
Loop Calibrator’s display.
To replace the battery:
1. Lift and remove the battery cover to gain access
to the battery compartment. See Figure 1.
2. Lift the battery from the case bottom and insert a
new 9 V battery.
3. Make sure the positive and negative battery posts
are oriented correctly, as noted in the bottom of
the battery compartment.
AAK07F .emf
Figure 1. Replacing the Battery
Disassembling and Reassembling the
Calibrator
W Warning
To prevent electric shock or personal
injury:
Do not allow water into the case.
Remove any input signals prior to
removing test leads and opening case.
When servicing the Loop Calibrator,
use only specified replacement parts.
To disassemble and reassemble the Loop Calibrator,
refer to Figures 2 and 3 as necessary and perform the
following procedure.
Note
Both a Phillips-head screwdriver and a flat-
blade screwdriver are required.
Removing the Printed Circuit Assembly (PCA)
1. Make sure the Loop Calibrator is off.
2. Remove all test leads from the terminals of the
Loop Calibrator.
3. Gently pry the rotary knob off with a small flat-
blade screwdriver.
4. Lift and remove the battery cover to gain access
to the battery compartment an se screws d 2 ca
inside. Remove the battery.
5. Remove all Phillips-head screws from the case
bottom.
6. Separate the case top from the case bottom.
Verify that the battery terminals are being guided
through the bottom case before completely
removing the case bottom.
7. R eferring to Figure 3, insert a small, flat-blade
screwdriver between the edge of the case top and
the PCA. Gently unsnap the case top from the
PCA.
8. Repeat step 7 for each of the remaining snaps that
secure the PCA to the case top.
9. Lift the PCA from the case top by its edges.
10. If the elastomeric strips for the keypad assembly
and LCD are stuck to the PCA, remove them, but
do not touch the conductive edges.


Specyfikacje produktu

Marka: Fluke
Kategoria: Niesklasyfikowane
Model: 707

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Fluke 707, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Niesklasyfikowane Fluke

Fluke

Fluke ii500 Instrukcja

9 Października 2024
Fluke

Fluke iSee TC01A Instrukcja

9 Października 2024
Fluke

Fluke 438-II Instrukcja

9 Października 2024
Fluke

Fluke T5-1000 Instrukcja

4 Października 2024
Fluke

Fluke TiX885 Instrukcja

29 Września 2024
Fluke

Fluke TiX875 Instrukcja

29 Września 2024
Fluke

Fluke TiX880 Instrukcja

29 Września 2024
Fluke

Fluke TiX870 Instrukcja

29 Września 2024
Fluke

Fluke 700HTH Instrukcja

29 Września 2024

Instrukcje Niesklasyfikowane

Najnowsze instrukcje dla Niesklasyfikowane