Instrukcja obsługi Fluke 323


Przeczytaj poniżej 📖 instrukcję obsługi w języku polskim dla Fluke 323 (2 stron) w kategorii Sprzęt pomiarowy. Ta instrukcja była pomocna dla 6 osób i została oceniona przez 2 użytkowników na średnio 4.5 gwiazdek

Strona 1/2
PN 4637502 March 2015
© 2015 Fluke Corporation. All rights reserved.
323/324/325
Clamp Meters
Quick Reference Guide
 Warning
To prevent possible electrical shock, fire, or personal
injury, read the “323/324/325 Safety Sheet” before you
use this Product.
30 0 V
CAT
60 0 V
CAT
30 0 VCAT
60 0 VCAT
325
TRUE RMS CLAMP METER
30 0 V
CAT
60 0 V
CAT
30 0 VCAT
60 0 VCAT
323
TRUE RMS CLAMP METER
30 0 V
CAT
60 0 V
CAT
30 0 VCAT
60 0 VCAT
324
TRUE RMS CLAMP METER
30 0 V
CAT
60 0 V
CAT
300 VCAT
600 VCAT
325
TRUE RMS CLAMP METER
V
CAT
60 0 V
CAT
300 VCAT
600 VCAT
325
TRUE RMS CLAMP METER
325
323, 324
300 VCAT
VV
or >30 V
V
323 V dc -
324, 325
V dc & V ac -
V
323 V ac -
20
OFF
ON V
Hz AC
DC
A
k
MAXMIN
F
FC
ON
323
324/325
ON
2:00
2:00
MIN
MAX
300 V
CAT
600 V
CAT
300 VCAT
600 VCAT
325
TRUE RMS CLAMP METER
300 V
CAT
600 V
CAT
300 VCAT
600 VCAT
325
TRUE RMS CLAMP METER
<70323
324/325 <30
325
324
323
Electrical Specifications
AC Current (Jaw)
Range 323 . . . . . . . . 400.0 A
324, 325 . . . . (40.00, 400.0) A
Resolution 323 . . . . . . . . 0.1 A
324, 325 . . . . (0.01, 0.1) A
Accuracy 323, 325 . . . . 2.0 % ±5 digits (45 – 65 Hz)
2.5 % ±5 digits (65 – 400 Hz)
324 . . . . . . . . 1.5 % ±5 digits (45 – 400 Hz)
Note: Add 2 % for position sensitivity.
DC Current with Jaw (325)
Range . . . . . . . . . . . (40.00, 400.0) A
Resolution . . . . . . . . . . . (0.01, 0.1) A
Accuracy . . . . . . . . . . . 2.0 % ± 5 digits
AC Voltage
Range . . . . . . . . . . . 600.0 V
Resolution . . . . . . . . . . . 0.1 V
Accuracy (45 – 400 Hz) . . 1.5 % ± 5 digits
DC Voltage
Range . . . . . . . . . . . 600.0 V
Resolution . . . . . . . . . . . 0.1 V
Accuracy . . . . . . . . . . . 1 % ±5 digits
Resistance
Range 323, 324 . . . . (400.0, 4000)
325 . . . . . . . . (400.0, 4000, 40000)
Resolution . . . . . . . . . . . (0.1, 1, 10)
Accuracy . . . . . . . . . . . 1 % ±5 digits
Continuity Beeper
323 . . . . . . . . 70 ≤ Ω
324/325 . . . . . 30 ≤ Ω
Capacitance (324, 325)
Range . . . . . . . . . . . (100.0, 1000) Fµ
Resolution . . . . . . . . . . . (0.1, 1) Fµ
Accuracy . . . . . . . . . . . 1 % ±4 digits
Frequency with Jaw (325)
Range . . . . . . . . . . . 5.0 to 500.0 Hz
Resolution . . . . . . . . . . . 0.1 Hz
Accuracy . . . . . . . . . . . 0.5 % ±4 digits
Trigger Level . . . . . . . . . . 5 to 10 Hz, 10 A
10 to 100 Hz, 5 A
100 to 500 Hz, 10 A
Specifications
Contact Temperature (324, 325)
Range . . . . . . . . . . . . . . -10.0 °C to 400.0 °C
Resolution . . . . . . . . . . . . . . 0.1 °C
Accuracy . . . . . . . . . . . . . . 1 % ±8 digits
Note: Temperature uncertainty (accuracy) does not include error of
the thermocouple probe.
Mechanical Specifications
Size (L x W x H) . . . . . . . . . . (207 x 75 x 34) mm
Weight
323 . . . . . . . . . . . 265 g
324 . . . . . . . . . . . 208 g
325 . . . . . . . . . . . 283 g
Environmental Specifications
Operating Temperature . . . . . -10 °C to +50 °C
Storage Temp . . . . . . . . . . . . -30 °C to +60 °C
Operating Humidity . . . . . . . . Non Condensing ( 10 °C)
90 % RH (at 10 °C to 30 °C)
75 % RH (at 30 °C to 40 °C)
45 % RH (at 40 °C to 50 °C)
(Without Condensation)
Operating Altitude . . . . . . . . . 2000 meters
Storage Altitude . . . . . . . . . . . 12,000 meters
EMI, EMC
Electromagnetic
Compatibility . . . . . . . . . . . . . Applies to use in Korea only. Class A
Equipment (Industrial Broadcasting &
Communication Equipment) [1]
[1] This product meets requirements for
industrial (Class A) electromagnetic
wave equipment and the seller or
user should take notice of it. This
equipment is intended for use in
business environments and is not to
beused in homes.
Temperature Coefficients . . . . Add 0.1 x specified accuracy for each
degree C above 28 °C or below 18 °C
Over Voltage Category . . . . . CAT IV 300 V, CAT III 600 V
Safety Compliance . . . . . . . . EN/IEC 61010-1, Pollution Degree 2
EN/IEC 61010-2-032
EN/IEC 61010-031:2002/A1:2008
Agency Approvals . . . . . . . . . Complies with CAN/CSA-C22.2
No. 61010-1, 2nd edition, including
Amendment 1.,  ,
IP Rating . . . . . . . . . . . . . . IP 30 Per IEC 60529:2001;
Non-operating
Batteries . . . . . . . . . . . . . . 2 AAA, NEDA 24A, IEC LR03
300 V
CAT
600 V
CAT
300 V
CAT
600 VCAT
325
TRUE RMS CLAMP MET ER
300VCAT
600VCAT
325
TRUE RMS CLAMP METER
Hz
Hz
324, 325 325
AAA
300VCAT
600VCAT
325
TRUE RMS CLAMP METER
324, 325
HOLD
20 V 30 V
>
300VCAT
600VCAT
325
TRUE RMS CLAMP METER
AC/DC & ZERO
AC/DC
ZERO
AC/DC
ZERO
AC/DC
ZERO 2 sec


Specyfikacje produktu

Marka: Fluke
Kategoria: Sprzęt pomiarowy
Model: 323

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Fluke 323, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Sprzęt pomiarowy Fluke

Fluke

Fluke 319 Instrukcja

24 Września 2024
Fluke

Fluke T150/VDE Instrukcja

23 Września 2024
Fluke

Fluke i400 Instrukcja

15 Września 2024
Fluke

Fluke 301D Instrukcja

15 Września 2024
Fluke

Fluke 1507 Instrukcja

7 Września 2024
Fluke

Fluke T110VDE Instrukcja

4 Września 2024
Fluke

Fluke i400s Instrukcja

4 Września 2024
Fluke

Fluke 1503 Instrukcja

3 Września 2024
Fluke

Fluke 393 Instrukcja

31 Sierpnia 2024

Instrukcje Sprzęt pomiarowy

Najnowsze instrukcje dla Sprzęt pomiarowy