Instrukcja obsługi CyberPower MPV615P


Przeczytaj poniżej 📖 instrukcję obsługi w języku polskim dla CyberPower MPV615P (2 stron) w kategorii Niesklasyfikowane. Ta instrukcja była pomocna dla 7 osób i została oceniona przez 2 użytkowników na średnio 4.5 gwiazdek

Strona 1/2
6-OUTLET MEDICAL GRADE
POWER STRIP
AND
6-OUTLET MEDICAL GRADE
SURGE PROTECTOR
MPV615P/MPV615S
USER MANUAL
YOUR ULTIMATE ALLY IN POWER
Cyber Power Systems (USA), Inc.
4241 12th Avenue East, Suite 400 | Shakopee, MN 55379
CyberPowerSystems.com
INSTALLATION INSTRUCTIONS
1. Use the screw hole on either end of the device for fixed
mounting.
OR: Use the keyhole slots for hanging on a mounting
surface. Mark the locations of the screw holes to align with
the keyhole slot centers.
(Note: The device will drop approximately 1/4” when it is
locked in place). Drill screw holes and insert screws until
the screw heads are 1/16” from the mounting surface.
Engage keyhole slots on the screw heads and gently push
the device downward to lock it in place.
2. Use a slotted screwdriver to remove the knurled nut (a)
on the return ground terminal (b) to allow insertion of
the user supplied ground conductor (c) from the patient
equipment grounding point (d). Refasten (hand-tighten
plus a quarter turn) the knurled nut to secure the ground
conductor in place.
a
b
c
d
e
f
e
3. Open the outlet covers (e). For safe operation, the outlet
covers do not open easily. If the covers are dicult to
open, use a slotted screwdriver to carefully open them by
pushing the cover release towards the center of the cover.
Insert the plugs of the authorized medical equipment (f).
Note: Open outlet covers only as needed. Close unused
outlet covers.
4. Power the outlet strip.
For MPV615P, the green LED indicator will be illuminated
when it is powered.
For MPV615S, the green LED indicator will be illuminated
when the surge protection feature is working properly.
Note (For MPV615S): This device features an internal protec-
tion that will disconnect the surge protective component at
the end of its useful life, but will maintain unprotected power
to the load. If this situation is undesirable for the application,
replace the device.
FOR MORE INFORMATION
Visit CyberPowerSystems.com for more information regarding:
• Product information and certifications
• Product warranty
• Connected equipment guarantee
TECHNICAL SUPPORT
Visit: CyberPowerSystems.com/support
Toll-Free: 1-877-297-6937
© 2022 Cyber Power Systems (USA), Inc. All rights reserved.
All other trademarks are the property of their respective owners.
CAUTION
TO REDUCE THE RISK OF ELECTRIC SHOCK:
Use only in dry locations and only indoors. Use only with duplex receptacle having
center screw.
DO NOT plug into another health care outlet assembly, a relocatable power tap, or
an extension cord. Connect only AUTHORIZED medical equipment verified for safe
touch and leakage current.
DO NOT daisy chain” surge protectors.
DO NOT use with any aquarium equipment.
DO NOT use if properly grounded outlets are not available.
DO NOT install this device if there is less than 10 meters (30 feet) of wire between
the electrical outlet and electrical service panel.
Not for Home Health Care Use.
WARNING: RISK OF DEATH OR SERIOUS INJURY. Only plug into a permanently
installed branch circuit receptacle and use within the Patient Care Vicinity only when
a protective earth grounding conductor also connects the Equipment Grounding
(Protective Earth) Terminal to the Patient Equipment Grounding Point.
WARNING: After verifying the combined touch and leakage current secure the outlet
covers closed.See installation instructions for tightening details.
WARNING: To reduce the risk of fire, electric shock, or injury to persons:
- Use this HCOA (Health Care Facility Outlet Assembly) only for its intended use as
described in these instructions. Do not use attachments not recommended by the
manufacturer.
- Never operate this HCOA if it has a damaged cord or plug, if it is not working prop-
erly, if it has been dropped or damaged, or dropped into water
- Keep the cord away from heated surfaces.
- Do not use outdoors.
- Do not place on floor.
- Mount and position to prevent any liquid from entering the HCOA.
WARNING: RISK OF ELECTRIC SHOCK - Connect this HCOA to a properly grounded
Hospital Grade (green dot) Receptacle only.
Grounding Instructions: This product must be grounded and, when used within a
Patient Care Vicinity, must also be grounded to the Patient Equipment Grounding
Point. If malfunction or insulation breakdown occurs, grounding provides a path of
least resistance for electric current to reduce the risk of electric shock. This product
is equipped with a cord having an equipment-grounding conductor and a grounding
plug. The plug must be plugged into an appropriate outlet that is properly installed
and grounded in accordance with all local codes and ordinances. This product is also
provided with a “Patient Equipment Grounding Terminal” (or jack) for connection of
a “Patient Equipment Grounding Conductor” which in turn must be connected to the
“Patient Equipment Grounding Point” before use within any “Patient Care Vicinity”
6ft (1.8m), beyond the patient bed, chair, table, treadmill, or other device that sup-
ports the patient during examination and treatment.
DANGER: Improper connection of the equipment-grounding conductor and Protec-
tive Earth Grounding Conductor can result in a risk of electric shock.
Check with a qualified electrician or service person if you are in doubt as to whether
the product is properly grounded.
Check to see that the HCOA rating is suitable for the application. The total true R.M.S.
current draw of the devices to be protected should not exceed 15 Amps (or 20 Amps
based on the model).
CONFORMANCE APPROVALS
This device complies with Part 15 of the FCC Rules.
Operation is subject to the following two conditions: (1) this
device may not cause harmful interference, and (2) this device must
accept any interference received, including interference that may
cause undesired operation.
This equipment has been tested and found to comply with the limits
for a Class B digital device, pursuant to Part 15 of the FCC Rules.
These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment
generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with the instructions, may cause
harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does cause harmful interfer-
ence to radio or television reception, which can be determined by
turning the equipment o and on, the user is encouraged to try to
correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit dierent from
that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
WARNING: Changes or modifications not expressly approved by the
party responsible for compliance could void the user’s authority to
operate the equipment.
YOUR ULTIMATE ALLY IN POWER
SPECIFICATIONS
Model Number: MPV615P
Outlets: Standard Grounded Outlets (6)
Electrical Rating: 125 V / 15 A / 1875 W
Power Cord Length: 15 ft
Model Number: MPV615S
Outlets: Standard Grounded Outlets (6)
Surge Protection: 1,560 Joules
Electrical Rating: 125 V / 15 A / 1875 W
Power Cord Length: 15 ft
UL Clamping Voltage: UL 1449 4th /600 V (L-N)
Maximum Peak Current: 45,000 A
AC Line Protected: L-N: 45,000 A
Response Time: Less than 1 nanosecond
i WARNING: This product can expose you to chemicals including bisphenol A (BPA) and
styrene (ABS), which is known to the State of California to cause reproductive harm and
cancer. For more information, go to www.P65Warnings.ca.gov.


Specyfikacje produktu

Marka: CyberPower
Kategoria: Niesklasyfikowane
Model: MPV615P

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z CyberPower MPV615P, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Niesklasyfikowane CyberPower

Instrukcje Niesklasyfikowane

Najnowsze instrukcje dla Niesklasyfikowane

Stihl

Stihl HT 70 Instrukcja

15 Października 2024
Ernesto

Ernesto H14270 Instrukcja

15 Października 2024
Uniden

Uniden BT MIC KIT Instrukcja

15 Października 2024
Vaude

Vaude Omnis Bike 26 Instrukcja

15 Października 2024
ART

ART ProChannel II Instrukcja

15 Października 2024
Aukey

Aukey DR02J Instrukcja

15 Października 2024
Master Lock

Master Lock 653EURD Instrukcja

15 Października 2024
Roland

Roland FR-18 Diatonic Instrukcja

15 Października 2024