Data Sheet
Solar Analyzer
Professional analyzer for testing, maintenance, troubleshooting
and efciency of solar panels. The analyzer is used in the
installation of solar panels to determine the proper inverter size,
optimum power output position of panels and identies broken
or worn-out cells.
SOLAR 600
I-V Curve test for solar cell
Maximum Solar Power (Pmax) search by auto-scan (60V, 12A)
Maximum Voltage (Vmaxp) at Pmax
Maximum Current (Imaxp) at Pmax
Voltage at open circuit (Vopen)
Current at short circuit (Ishort)
I-V curve with cursor
Calculation of panel efciency (%)
Manual single point test
Real Time data logging
PC Download
Amprobe® Test Tools www.Amprobe.com
Data Sheet
General Specications
Battery Type Rechargeable Lithium Battery, 11.1Vdc, 3400mAh
Battery Life 400 times of linear scan from 60V to 0V and 0A to 12A
Data Logging Memory Size 99 records
AC Adaptor AC 110V or 220V input DC 15V / 1~3A output
Dimension 257(L) x 155(W) x 57(H) mm (10.12 x 6.1 x 2.24 inch)
Weight 1160g / 40.0oz (Batteries included)
Operation Environment C to 50°C (32°F to 122°F), 85% RH
Temperature Coefcient 0.1% of full scale / °C (<18°C or >28°C)
Storage Temperature -20°C to 60°C (-F to 140°F), 75% RH
SOLAR 600 Solar Analyzer
Amprobe® Test Tools www.Amprobe.com
Data Sheet
Electrical Specications
Range Accuracy
DC Voltage 0.0009.999V ±1% ± (1% of Vopen ± 0.1V)
10.0060.00V
DC Current 0.01–9.99A ±1% ± (1% of Ishort ± 9mA)
10.0–12.0A ±1% ± (1% of Ishort ± 0.09A)
DC Current Simulation 0.01 ~ 9.999 A ± 1 % ± 9 mA
10 ~ 12 A ± 1 % ± 0.09A
If current is greater than 12A, test (Auto-Scan, Scan, or Test) can not be performed.
Included Accessories Soft carrying case, Users manual, AC Adaptor, RS232C (to USB Bridge) Cable, Rechargeable Lithium
Battery Pack, Software CD, Software manual, Kelvin Clips (12A max, 1 pair)
SOLAR 600 Solar Analyzer


Specyfikacje produktu

Marka: Amprobe
Kategoria: Niesklasyfikowane
Model: SOLAR-600

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Amprobe SOLAR-600, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Niesklasyfikowane Amprobe

Amprobe

Amprobe AC40C Instrukcja

3 Października 2024
Amprobe

Amprobe DM-CT-100 Instrukcja

30 Września 2024
Amprobe

Amprobe AD40B Instrukcja

29 Września 2024
Amprobe

Amprobe WT-10 Instrukcja

27 Września 2024
Amprobe

Amprobe TMD-51 Instrukcja

23 Września 2024
Amprobe

Amprobe THWD-10W Instrukcja

16 Września 2024
Amprobe

Amprobe NCV-1020 Instrukcja

15 Września 2024
Amprobe

Amprobe WT-60 Instrukcja

15 Września 2024
Amprobe

Amprobe CT238A Instrukcja

13 Września 2024
Amprobe

Amprobe WT-20 Instrukcja

13 Września 2024

Instrukcje Niesklasyfikowane

Najnowsze instrukcje dla Niesklasyfikowane