INSULATION TESTER
1. SAFETY INFORMATION
The circuit under test must be de-energized and isolated before
connections are made except for voltage measurement.
Verify operation prior to measuring hazardous voltages (voltage
above 30V AC rms, 42V AC peak and 60V DC).
AMB-25
Do not touch the Circuit connections during a test.
Disconnect the live test lead before disconnecting the neutral
test lead.
INSTRUCTION MANUAL
After insulation tests, to protect electric shock, capacitive circuits
must to be discharged.
READ
MEM
Caution refer to this manual before using the meter.
Dangerous voltages.
Meter is protected throughout by double insulation or
reinforced insulation.
Comply with IEC1010-1
When servicing, use only specified replacement parts.
s
AC
m:
HOLD
VG
K
M
Do not use the meter if the low battery indicator (BT ) is
showed.
Test leads (including crocodile clips) must be in good order,
clean and no broken or cracked.
Do not use the meter if it looks damaged.
Do not use the meter around explosive gas, vapor or dust.
Do not push test button before all connection and preparation is
done. The instrument must only be used by suitably trained and
competent persons.
Do not use the meter with any parts or cover removed.
Do not use the meter in a wet environment.
U.S. PAT. NO. 478,017
JAPAN PAT. NO. 1180870
CHINA PAT. NO. ZL02367250.1
Warnings and Safety symbols:
1
2-2 Electrical Specifications
2. SPECIFICATIONS Accuracies are specified as:
±(...% of reading + ...digits) at 23
±5,below 80% RH.
2-1 General Information
Environment conditions : Insulation Resistance
c Installation Categories 1000V Range Resolution Accuracy Test Voltages
40MΩ/400MΩ 250V+20% ~ -0%
400MΩ/1000MΩ
/500V 500V+20% ~ -0%
400M
Ω/ 2000MΩ
/1000V
40MΩ: 10 KΩ
400MΩ:100 KΩ
4000MΩ: 10M
Ω
3%+5
(<1000M)
5%+5
(>1000M) 1000V+20% ~ -0%
Analog Bar Graph 0 to 4000M
Ω
Nominal Current 1mA
d
Pollution Degree 2
e Altitude up to 2000 meters
f Indoor use only
Safety Meets of IEC61010-1 and IEC61557
Display : Dual display, 3-3/4 Digital readout with analog bar
indication.
Sampling Rate : 2.5 sample/sec.
Manual data Memory and Read : Memory capacity 9 set.
Over Range Indicator : ” will be displayed.
Low Battery Indication :
BT Resistance & Continuity (<40
Ω
Ω
Ω
ΩΩ )
The ( ) will be displayed when the battery voltage drop
below the operating voltage.
Range Resolution Accuracy Max. open Circuit
Voltage
Overload
Protection
999.9Ω0.1Ω 1%+2 3V 1000Vrms
Operating Temperature and Humidity
0 to 50 (32 to 122) below 80% RH (noncondensing).
Temperature Coefficient : 0.10 x (specified accuracy)/
Storage Temperature and Humidity :
-10 to 60 (14 to 140
) below 70% RH (noncondensing)
Battery : 6 x 1.5V Size “AA” battery.
Fuse : 6mm x 32mm (0.25 x 1.25 inch), 0.5A 1000V, Fast Acting.
Dimensions : 235 (L) x 116 (W) x 54(H) mm ,
(9.3”L x 4.6”W x 2.1”H)
Weight : Approx. 520g (1.15 LB), include battery
Accessories : Test leads, 6pcs battery, Holster, operation manual.
3
2
c LCD display.
3. PARTS & CONTROLS
COM V
d MEM key : Manual data memory control key.
MEM
READ
M
V
V M
e LOCK key : Locks the test Insulation Resistance (M
Ω).
a). Press and hold down TEST key, then press LOCK key, the
LOCK icon appears on the display.
b). In MΩ mode continuously applies the test voltage to the circuit to
be tested. The beeper sounds every 2 seconds to remind you
that you are in the LOCK mode.
c). Press Lock or TEST key again to exit the mode.
f READ key: Manual memory data reading control key.
g key: In Ω function, turns the test lead resistance
compensation ON.
Touch the probe tips together, then press key,
the icon appears on the display and the main
display indicates 0.0
Ω.
h TEST key :
a). Used for MΩ test functions.
b). Press and hold TEST key until the main display reading is stable.
i j ▼▲ keys : In READ mode, select the memory data location for
direct reading the data from the display.
k Rotary switch : To select a measurement function.
l Input terminals.
5
4


Specyfikacje produktu

Marka: Amprobe
Kategoria: Niesklasyfikowane
Model: AMB-25

Potrzebujesz pomocy?

Jeśli potrzebujesz pomocy z Amprobe AMB-25, zadaj pytanie poniżej, a inni użytkownicy Ci odpowiedzą




Instrukcje Niesklasyfikowane Amprobe

Amprobe

Amprobe AC40C Instrukcja

3 Października 2024
Amprobe

Amprobe DM-CT-100 Instrukcja

30 Września 2024
Amprobe

Amprobe AD40B Instrukcja

29 Września 2024
Amprobe

Amprobe WT-10 Instrukcja

27 Września 2024
Amprobe

Amprobe TMD-51 Instrukcja

23 Września 2024
Amprobe

Amprobe THWD-10W Instrukcja

16 Września 2024
Amprobe

Amprobe NCV-1020 Instrukcja

15 Września 2024
Amprobe

Amprobe WT-60 Instrukcja

15 Września 2024
Amprobe

Amprobe CT238A Instrukcja

13 Września 2024
Amprobe

Amprobe WT-20 Instrukcja

13 Września 2024

Instrukcje Niesklasyfikowane

Najnowsze instrukcje dla Niesklasyfikowane

Stihl

Stihl HT 70 Instrukcja

15 Października 2024
Ernesto

Ernesto H14270 Instrukcja

15 Października 2024
Uniden

Uniden BT MIC KIT Instrukcja

15 Października 2024
Vaude

Vaude Omnis Bike 26 Instrukcja

15 Października 2024
ART

ART ProChannel II Instrukcja

15 Października 2024
Aukey

Aukey DR02J Instrukcja

15 Października 2024
Master Lock

Master Lock 653EURD Instrukcja

15 Października 2024
Roland

Roland FR-18 Diatonic Instrukcja

15 Października 2024